» Articles » PMID: 29520182

Self-Gated Late Gadolinium Enhancement at 7T to Image Rats with Reperfused Acute Myocardial Infarction

Overview
Journal Korean J Radiol
Specialty Radiology
Date 2018 Mar 10
PMID 29520182
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: A failed electrocardiography (ECG)-trigger often leads to a long acquisition time (TA) and deterioration in image quality. The purpose of this study was to evaluate and optimize the technique of self-gated (SG) cardiovascular magnetic resonance (CMR) for cardiac late gadolinium enhancement (LGE) imaging of rats with myocardial infarction/reperfusion.

Materials And Methods: Cardiovascular magnetic resonance images of 10 rats were obtained using SG-LGE or ECG with respiration double-gating (ECG-RESP-gating) method at 7T to compare differences in image interference and TA between the two methods. A variety of flip angles (FA: 10°-80°) and the number of repetitions (NR: 40, 80, 150, and 300) were investigated to determine optimal scan parameters of SG-LGE technique based on image quality score and contrast-to-noise ratio (CNR).

Results: Self-gated late gadolinium enhancement allowed successful scan in 10 (100%) rats. However, only 4 (40%) rats were successfully scanned with the ECG-RESP-gating method. TAs with SG-LGE varied depending on NR used (TA: 41, 82, 154, and 307 seconds, corresponding to NR of 40, 80, 150, and 300, respectively). For the ECG-RESP-gating method, the average TA was 220 seconds. For SG-LGE images, CNR (42.5 ± 5.5, 43.5 ± 7.5, 54 ± 9, 59.5 ± 8.5, 56 ± 13, 54 ± 8, and 41 ± 9) and image quality score (1.85 ± 0.75, 2.20 ± 0.83, 2.85 ± 0.37, 3.85 ± 0.52, 2.8 ± 0.51, 2.45 ± 0.76, and 1.95 ± 0.60) were achieved with different FAs (10°, 15°, 20°, 25°, 30°, 35°, and 40°, respectively). Optimal FAs of 20°-30° and NR of 80 were recommended.

Conclusion: Self-gated technique can improve image quality of LGE without irregular ECG or respiration gating. Therefore, SG-LGE can be used an alternative method of ECG-RESP-gating.

Citing Articles

Emerging methods for the characterization of ischemic heart disease: ultrafast Doppler angiography, micro-CT, photon-counting CT, novel MRI and PET techniques, and artificial intelligence.

Willemink M, Varga-Szemes A, Schoepf U, Codari M, Nieman K, Fleischmann D Eur Radiol Exp. 2021; 5(1):12.

PMID: 33763754 PMC: 7991013. DOI: 10.1186/s41747-021-00207-3.


Guidelines for Cardiovascular Magnetic Resonance Imaging from the Korean Society of Cardiovascular Imaging-Part 2: Interpretation of Cine, Flow, and Angiography Data.

Lee J, Hur J, Yang D, Lee B, Im D, Hong S Korean J Radiol. 2019; 20(11):1477-1490.

PMID: 31606953 PMC: 6791819. DOI: 10.3348/kjr.2019.0407.


Guideline for Cardiovascular Magnetic Resonance Imaging from the Korean Society of Cardiovascular Imaging-Part 1: Standardized Protocol.

Jo Y, Kim J, Park C, Lee J, Hur J, Yang D Korean J Radiol. 2019; 20(9):1313-1333.

PMID: 31464111 PMC: 6715561. DOI: 10.3348/kjr.2019.0398.

References
1.
Brinegar C, Wu Y, Foley L, Hitchens T, Ye Q, Ho C . Real-time cardiac MRI without triggering, gating, or breath holding. Annu Int Conf IEEE Eng Med Biol Soc. 2009; 2008:3381-4. PMC: 2793087. DOI: 10.1109/IEMBS.2008.4649931. View

2.
Tham E, Hung R, Myers K, Crawley C, Noga M . Optimization of myocardial nulling in pediatric cardiac MRI. Pediatr Radiol. 2011; 42(4):431-9. DOI: 10.1007/s00247-011-2276-z. View

3.
Klein C, Schmal T, Nekolla S, Schnackenburg B, Fleck E, Nagel E . Mechanism of late gadolinium enhancement in patients with acute myocardial infarction. J Cardiovasc Magn Reson. 2007; 9(4):653-8. DOI: 10.1080/10976640601105614. View

4.
Vallee J, Ivancevic M, Nguyen D, Morel D, Jaconi M . Current status of cardiac MRI in small animals. MAGMA. 2004; 17(3-6):149-56. DOI: 10.1007/s10334-004-0066-4. View

5.
Protti A, Sirker A, Shah A, Botnar R . Late gadolinium enhancement of acute myocardial infarction in mice at 7T: cine-FLASH versus inversion recovery. J Magn Reson Imaging. 2010; 32(4):878-86. DOI: 10.1002/jmri.22325. View