Bony Ingrowth Potential of 3D-printed Porous Titanium Alloy: a Direct Comparison of Interbody Cage Materials in an in Vivo Ovine Lumbar Fusion Model

Overview
Journal Spine J
Specialty Orthopedics
Date 2018 Mar 3
PMID 29496624
Citations 82
Authors
Affiliations
Soon will be listed here.
Abstract

Background Context: There is significant variability in the materials commonly used for interbody cages in spine surgery. It is theorized that three-dimensional (3D)-printed interbody cages using porous titanium material can provide more consistent bone ingrowth and biological fixation.

Purpose: The purpose of this study was to provide an evidence-based approach to decision-making regarding interbody materials for spinal fusion.

Study Design: A comparative animal study was performed.

Methods: A skeletally mature ovine lumbar fusion model was used for this study. Interbody fusions were performed at L2-L3 and L4-L5 in 27 mature sheep using three different interbody cages (ie, polyetheretherketone [PEEK], plasma sprayed porous titanium-coated PEEK [PSP], and 3D-printed porous titanium alloy cage [PTA]). Non-destructive kinematic testing was performed in the three primary directions of motion. The specimens were then analyzed using micro-computed tomography (µ-CT); quantitative measures of the bony fusion were performed. Histomorphometric analyses were also performed in the sagittal plane through the interbody device. Outcome parameters were compared between cage designs and time points.

Results: Flexion-extension range of motion (ROM) was statistically reduced for the PTA group compared with the PEEK cages at 16 weeks (p-value=.02). Only the PTA cages demonstrated a statistically significant decrease in ROM and increase in stiffness across all three loading directions between the 8-week and 16-week sacrifice time points (p-value≤.01). Micro-CT data demonstrated significantly greater total bone volume within the graft window for the PTA cages at both 8 weeks and 16 weeks compared with the PEEK cages (p-value<.01).

Conclusions: A direct comparison of interbody implants demonstrates significant and measurable differences in biomechanical, µ-CT, and histologic performance in an ovine model. The 3D-printed porous titanium interbody cage resulted in statistically significant reductions in ROM, increases in the bone ingrowth profile, as well as average construct stiffness compared with PEEK and PSP.

Citing Articles

Clinical and radiological outcomes of titanium cage versus polyetheretherketone cage in lumbar interbody fusion: a systematic review and meta-analysis.

Wang H, Zhang H, Xiao C, Zhang K, Qi L Neurosurg Rev. 2025; 48(1):295.

PMID: 40075000 DOI: 10.1007/s10143-025-03453-w.


Biomechanical analysis of a newly designed and 3D printed plate-locking interbody cage: an observational study of finite element analysis.

Ni S, Yang R, Liu S, Hu Y Sci Rep. 2025; 15(1):3534.

PMID: 39875489 PMC: 11775238. DOI: 10.1038/s41598-025-88151-9.


Artificial Intelligence-Guided Inverse Design of Deployable Thermo-Metamaterial Implants.

Jiao P, Zhang C, Meng W, Wang J, Jang D, Wu Z ACS Appl Mater Interfaces. 2025; 17(2):2991-3001.

PMID: 39746033 PMC: 11744508. DOI: 10.1021/acsami.4c17625.


Clinical Application of 3D-Printed Artificial Vertebral Body (3DP AVB): A Review.

Kiselev R, Zheravin A J Pers Med. 2024; 14(10).

PMID: 39452532 PMC: 11508315. DOI: 10.3390/jpm14101024.


A novel porous interbody fusion cage modified by microarc oxidation and hydrothermal treatment technology accelerate osseointegration and spinal fusion in sheep.

Sun J, Liu S, Zou D, Ni R, Wei C, Wang H RSC Adv. 2024; 14(44):31966-31978.

PMID: 39391624 PMC: 11462409. DOI: 10.1039/d3ra08185k.


References
1.
De Leonardis D, Garg A, Pecora G . Osseointegration of rough acid-etched titanium implants: 5-year follow-up of 100 minimatic implants. Int J Oral Maxillofac Implants. 1999; 14(3):384-91. View

2.
Cunningham B, Orbegoso C, Dmitriev A, Hallab N, Sefter J, McAfee P . The effect of titanium particulate on development and maintenance of a posterolateral spinal arthrodesis: an in vivo rabbit model. Spine (Phila Pa 1976). 2003; 27(18):1971-81. DOI: 10.1097/00007632-200209150-00004. View

3.
Cunningham B, Orbegoso C, Dmitriev A, Hallab N, Sefter J, Asdourian P . The effect of spinal instrumentation particulate wear debris. an in vivo rabbit model and applied clinical study of retrieved instrumentation cases. Spine J. 2003; 3(1):19-32. DOI: 10.1016/s1529-9430(02)00443-6. View

4.
Ha S, Kirch M, Birchler F, Eckert K, Mayer J, Wintermantel E . Surface activation of polyetheretherketone (PEEK) and formation of calcium phosphate coatings by precipitation. J Mater Sci Mater Med. 2004; 8(11):683-90. DOI: 10.1023/a:1018535923173. View

5.
Yu S, Hariram K, Kumar R, Cheang P, Aik K . In vitro apatite formation and its growth kinetics on hydroxyapatite/polyetheretherketone biocomposites. Biomaterials. 2004; 26(15):2343-52. DOI: 10.1016/j.biomaterials.2004.07.028. View