» Articles » PMID: 29363770

Assessment of Image Quality and Scatter and Leakage Radiation of an Integrated MR-LINAC System

Overview
Journal Med Phys
Specialty Biophysics
Date 2018 Jan 25
PMID 29363770
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To assess the image quality, scatter, and leakage radiation of an integrated magnetic resonance linear accelerator (MR-LINAC or MRL) system.

Methods: A large American College of Radiology (ACR) magnetic resonance imaging (MRI) accreditation phantom was used to evaluate the MRI capabilities of the integrated MRL system compared with those of other diagnostic MRI systems. Multiple sets of T1 and T2/PD images were acquired with the linear accelerator positioned at various angles and with the radiation beam on and off. Images also were acquired on three different occasions over a period of about 12 months. Scatter and leakage radiation were measured with a large (150 cm ) ion chamber recalibrated for MV energy. For scatter measurements, a 25-cm stack of solid-water materials was placed at the isocenter on the patient couch to simulate a patient. The head leakage was measured at 1 m from the linear accelerator head in directions determined to produce the maximum leakage. All measurements were repeated with the magnetic field turned off to study the effects of the magnetic field.

Results: The geometric distortion, slice thickness accuracy, image uniformity, ghosting ratio, and high-contrast detectability were comparable to other 1.5 T diagnostic MRI scanners. No observable changes in image quality and no appreciable differences were found between radiation beam-on and beam-off images. The measured leakage and scattered radiation changed by less than 5% when the magnetic field was on compared to measurements with the field off. The beam stopper leakage was approximately 0.3 R/1000 MU, and because there was no direct beam imparted on the walls, a vault designed for a modern-day LINAC should have enough required radiation shielding to house the MRL.

Conclusions: The image quality generated by the MRI system of the integrated MRL was similar to that of a diagnostic MRI scanner. Interference from the MV radiation was minimal, and there was no measurable difference in image quality with the beam on and off. Scatter radiation and leakage radiation of the MRL system were within the expected range of a comparable MV-LINAC.

Citing Articles

Development of shielding evaluation and management program for O-ring type linear accelerators.

Choi D, Ahn S, Kim D, Choi S, Ahn W, Kim J Sci Rep. 2024; 14(1):10719.

PMID: 38729975 PMC: 11087655. DOI: 10.1038/s41598-024-60362-6.


Evaluation of MU2net as an online secondary dose check for MR guided radiation therapy with the Elekta unity MR linac.

Shoobridge A, Baines J Phys Eng Sci Med. 2022; 45(2):429-441.

PMID: 35380366 DOI: 10.1007/s13246-021-01085-0.


Longitudinal assessment of quality assurance measurements in a 1.5 T MR-linac: Part II-Magnetic resonance imaging.

Subashi E, Dresner A, Tyagi N J Appl Clin Med Phys. 2022; 23(6):e13586.

PMID: 35332990 PMC: 9398228. DOI: 10.1002/acm2.13586.


A Comparison of the Distortion in the Same Field MRI and MR-Linac System With a 3D Printed Phantom.

Liu X, Li Z, Rong Y, Cao M, Li H, Jia C Front Oncol. 2021; 11:579451.

PMID: 34150605 PMC: 8209415. DOI: 10.3389/fonc.2021.579451.


Stability of MRI contrast agents in high-energy radiation of a 1.5T MR-Linac.

Wang J, Salzillo T, Jiang Y, Mackeyev Y, Fuller C, Chung C Radiother Oncol. 2021; 161:55-64.

PMID: 34089753 PMC: 8324543. DOI: 10.1016/j.radonc.2021.05.023.