Precision of a New Ocular Biometer in Children and Comparison with IOLMaster

Overview
Journal Sci Rep
Specialty Science
Date 2018 Jan 24
PMID 29358637
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

To assess the repeatability and reproducibility of AL-Scan in agreement with those by the IOLMaster in healthy children, two skilled operators measured ocular parameters in 58 children. The parameters included keratometry (K) values, anterior chamber depth (ACD), axial length (AL), central corneal thickness (CCT), pupil diameter (PD), and corneal diameter (CD). The cohort comprised of 32 boys and 26 girls. The AL-Scan measurements showed high repeatability, as the test-retest repeatability (TRT) values of AL, CCT, ACD, Kf, Ks, Km, CD, and PD were 0.09 mm, 5.1 μm, 0.04 mm, 0.28 D, 0.24 D, 0.21 D, 0.39 mm, and 0.22 mm, respectively. The within-subject coefficient of variation (CoV) was low (<0.35%) and the intraclass correlation coefficients (ICC) of all parameters were >0.85. The interobserver reproducibility was excellent with low values of TRT and ICC > 0.95. The CoV of AL, CCT, ACD, and K was <0.22%. The 95% limits of agreement between the AL-Scan and the IOLMaster were narrow for all parameters except for CD. The repeatability and reproducibility of the new biometer, Al-Scan, was excellent for all parameters and can be routinely used in children to measure the biometric values.

Citing Articles

Real-world outcomes of Defocus Incorporated Multiple Segments lenses on retarding axial elongation in myopic children and adolescents.

Chun R, Wong K, Lam C, To C, Liu K, Wong Y Front Med (Lausanne). 2025; 11:1416286.

PMID: 39927269 PMC: 11803585. DOI: 10.3389/fmed.2024.1416286.


Agreement between a new fully automatic ocular biometer based on optical low-coherence reflectometry and an optical biometer based on Scheimpflug imaging combined with partial coherence interferometry.

Ning R, Xu H, Li Z, Yu J, Xu S, Lei C BMC Ophthalmol. 2024; 24(1):455.

PMID: 39420260 PMC: 11484231. DOI: 10.1186/s12886-024-03716-y.


Efficacy of orthokeratology lens with the modified small treatment zone on myopia progression and visual quality: a randomized clinical trial.

Gong G, Zhang B, Guo T, Liu G, Zhang J, Zhang X Eye Vis (Lond). 2024; 11(1):35.

PMID: 39218909 PMC: 11367740. DOI: 10.1186/s40662-024-00403-3.


Refraction and ocular biometric parameters in 3-to 6-year-old preschool children : a large-scale population-based study in Chengdu, China.

Mu J, Zhang Z, Wu X, Chen S, Geng H, Duan J BMC Ophthalmol. 2024; 24(1):207.

PMID: 38711043 PMC: 11071229. DOI: 10.1186/s12886-024-03467-w.


Comparison of the Accuracy of Seven Vault Prediction Formulae for Implantable Collamer Lens Implantation.

Wu H, Luo D, Chen J, Wang H, Zhong D Ophthalmol Ther. 2023; 13(1):237-249.

PMID: 37943482 PMC: 10776513. DOI: 10.1007/s40123-023-00844-4.


References
1.
Huang J, Savini G, Li J, Lu W, Wu F, Wang J . Evaluation of a new optical biometry device for measurements of ocular components and its comparison with IOLMaster. Br J Ophthalmol. 2014; 98(9):1277-81. DOI: 10.1136/bjophthalmol-2014-305150. View

2.
RETZLAFF J, Sanders D, Kraff M . Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990; 16(3):333-40. DOI: 10.1016/s0886-3350(13)80705-5. View

3.
Savini G, Carbonelli M, Barboni P, Hoffer K . Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011; 37(2):302-9. DOI: 10.1016/j.jcrs.2010.07.039. View

4.
Kola M, Duran H, Turk A, Mollamehmetoglu S, Kalkisim A, Erdol H . Evaluation of the Repeatability and the Reproducibility of AL-Scan Measurements Obtained by Residents. J Ophthalmol. 2014; 2014:739652. PMC: 4129666. DOI: 10.1155/2014/739652. View

5.
Holzer M, Mamusa M, Auffarth G . Accuracy of a new partial coherence interferometry analyser for biometric measurements. Br J Ophthalmol. 2009; 93(6):807-10. DOI: 10.1136/bjo.2008.152736. View