Observational Study in Healthy Volunteers to Define Interobserver Reliability of Ultrasound Haemodynamic Monitoring Techniques Performed by Trainee Doctors

Overview
Journal Eur J Emerg Med
Specialty Emergency Medicine
Date 2018 Jan 10
PMID 29315095
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: Bedside ultrasound is increasingly being used to guide fluid management in shocked patients. Little data exist on the inter-rater reliability of techniques used, especially when performed by nonexpert trainee doctors. The primary aim of this study is to measure the inter-rater reliability of five ultrasound techniques commonly used to guide fluid management: inferior vena cava collapsibility index (IVCCI), transthoracic echocardiography (TTE)-derived stroke volumes, ultrasound cardiac output monitor (USCOM) derived stroke volume and carotid artery blood flow and corrected flow time measurements.

Methods: Two Royal College of Emergency Medicine level one ultrasound-certified emergency medicine trainees performed paired ultrasound measurements on 31 healthy nonpatient volunteers. Inter-rater reliability was assessed through three indices: interclass correlation coefficient (ICC), limits of agreements (LOAs) derived from Band-Altman plots and the proportion of paired scans with absolute differences of less that 15% (defined as agreement).

Results: TTE-derived measurements performed the best overall, with an LOA of 22%, an ICC of 0.55 and an agreement of 80%. USCOM also performed well, with an LOA of 33%, an ICC of 0.68 and an agreement of 58%. IVCCI and carotid artery-derived measurements performed poorly across all indices.

Conclusion: TTE-derived measurements showed the highest level of inter-rater reliability and can thus be expected to provide reliable measures over time with different sonographer clinicians. USCOM interobserver reliability was also adequate for clinical use. However, on the basis of inter-reliability measures, IVCCI and carotid artery measurements were found to be inadequate for clinical use.

Citing Articles

A nursing-led sepsis response team guiding resuscitation with point-of-care ultrasound: A review and model for improving bundle compliance while individualizing sepsis care.

Nunnally J, Ko S, Ugale K, Lowe T, Bond J, Kenny J SAGE Open Med. 2024; 12:20503121241290378.

PMID: 39526100 PMC: 11549707. DOI: 10.1177/20503121241290378.


The correlation between carotid artery Doppler and stroke volume during central blood volume loss and resuscitation.

Kerrebijn I, Atwi S, Elfarnawany M, Eibl A, Eibl J, Taylor J Acute Crit Care. 2024; 39(1):162-168.

PMID: 38476069 PMC: 11002613. DOI: 10.4266/acc.2023.01095.


The Correlation between Carotid Artery Corrected Flow Time and Velocity Time Integral during Central Blood Volume Loss and Resuscitation.

Kerrebijn I, Munding C, Horner C, Atwi S, Elfarnawany M, Eibl A J Med Ultrasound. 2024; 31(4):309-313.

PMID: 38264586 PMC: 10802870. DOI: 10.4103/jmu.jmu_80_23.


Inferior Vena Caval Measures Do Not Correlate with Carotid Artery Corrected Flow Time Change Measured Using a Wireless Doppler Patch in Healthy Volunteers.

Kenny J, Prager R, Rola P, McCulloch G, Atwi S, Munding C Diagnostics (Basel). 2023; 13(23).

PMID: 38066832 PMC: 10706625. DOI: 10.3390/diagnostics13233591.


Reference ranges for third-trimester maternal cardiovascular function parameters measured in normotensive pregnant women using a non-invasive cardiac output monitor: A study based on data from the prospective PEACH cohort study.

Lihme F, Basit S, Persson L, Larsen M, Lauridsen K, Lykke J BJOG. 2023; 131(4):463-471.

PMID: 37735094 PMC: 10938999. DOI: 10.1111/1471-0528.17667.