Does Size Matter? Cryoballoon-Based Pulmonary Vein Isolation Using a Novel 25-mm Circular Mapping Catheter
Overview
Authors
Affiliations
Background: Real-time recording of pulmonary vein isolation (PVI) using a circular mapping catheter has become a key aspect of cryoballoon (CB) ablation. The aim of this study was to investigate the procedural safety, efficacy and rate of real-time pulmonary vein (PV) recording using a novel circular mapping catheter with a 25-mm loop size for CB-based PVI.
Methods and results: A total of 40 patients with symptomatic atrial fibrillation (AF) underwent PVI using a second-generation CB and a novel 25-mm circular mapping catheter. A total of 159 PV were identified and successfully isolated. Real-time PV recording was achieved in 80% of the PV. In 3 right inferior PV the circular mapping catheter had to be exchanged for a stiff guidewire due to insufficient mechanical support. Therefore, acute PVI using exclusively the circular mapping catheter was achieved in 156/159 PV (98%). Mean procedure and fluoroscopy times were 66±21 min and 15±6 min, respectively. Transient phrenic nerve palsy occurred in 1 patient as the only procedural complication.
Conclusions: The exclusive use of a novel 25-mm circular mapping catheter for CB ablation of AF results in a real-time PV recording rate of 80% and isolation of 98% of targeted PV.
Matsuda Y, Masuda M, Asai M, Iida O, Okamoto S, Ishihara T J Arrhythm. 2019; 35(3):535-541.
PMID: 31293705 PMC: 6595353. DOI: 10.1002/joa3.12192.
Mugnai G, Manfrin M, de Asmundis C, Stroker E, Longobardi M, Rauhe W Indian Pacing Electrophysiol J. 2019; 19(6):211-215.
PMID: 31238123 PMC: 6904823. DOI: 10.1016/j.ipej.2019.06.004.
Pott A, Baumhardt M, Weinmann K, Rattka M, Stephan T, Kessler M Int J Cardiol Heart Vasc. 2019; 23:100353.
PMID: 31065585 PMC: 6495074. DOI: 10.1016/j.ijcha.2019.100353.
Pott A, Petscher K, Baumhardt M, Stephan T, Rattka M, Paliskyte R Heart Vessels. 2018; 34(3):496-502.
PMID: 30302548 PMC: 6373539. DOI: 10.1007/s00380-018-1254-x.