Trajectory of Instantaneous Axis of Rotation in Fixed Lumbar Spine with Instrumentation

Overview
Publisher Biomed Central
Specialty Orthopedics
Date 2017 Nov 18
PMID 29145877
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Several studies showed instantaneous axis of rotation (IAR) in the intact spine. However, there has been no report on the trajectory of the IAR of a damaged spine or that of a fixed spine with instrumentation. It is the aim of this study to investigate the trajectory of the IAR of the lumbar spine using the vertebra of deer.

Methods: Functional spinal units (L5-6) from five deer were evaluated with six-axis material testing machine. As specimen models, we prepared a normal model, a damaged model, and a pedicle screw (PS) model. We measured the IAR during bending in the coronal and sagittal planes and axial rotation. In the bending test, four directions were measured: anterior, posterior, right, and left. In the rotation test, two directions were measured: right and left.

Results: The IAR of the normal model during bending moved in the bending direction. The IAR of the damaged model during bending moved in the bending direction, but the magnitude of displacement was bigger compared to that of the normal model. In the PS model, the IAR during bending test hardly moved. During rotation test, the IAR of the normal model and PS model located in the spinal canal, but the IAR of the damaged model located in the posterior part of the vertebral body.

Conclusions: In this study, the IAR of damaged model was scattering and that of PS model was concentrating. This suggests that higher mechanical load applied to the dura tube and nerve roots in the damaged model and less mechanical load applied to that in the PS model.

Citing Articles

ICR in human cadaveric specimens: An essential parameter to consider in a new lumbar disc prosthesis design.

Vanaclocha-Saiz A, Atienza C, Vanaclocha V, Belloch V, Santabarbara J, Jorda-Gomez P N Am Spine Soc J. 2022; 2:100016.

PMID: 35141586 PMC: 8820058. DOI: 10.1016/j.xnsj.2020.100016.


AutoBend: An Automated Approach for Estimating Intervertebral Joint Function from Bone-Only Digital Models.

Jones K, Brocklehurst R, Pierce S Integr Org Biol. 2021; 3(1):obab026.

PMID: 34661062 PMC: 8514422. DOI: 10.1093/iob/obab026.


The most appropriate titanium mesh cage size for anterior spinal reconstruction after single-level lumbar total en bloc spondylectomy: a finite element analysis and cadaveric validation study.

Paholpak P, Sirichativapee W, Wisanuyotin T, Kosuwon W, Kasai Y, Murakami H J Orthop Surg Res. 2021; 16(1):178.

PMID: 33750424 PMC: 7941739. DOI: 10.1186/s13018-021-02326-4.

References
1.
Wachowski M, Mansour M, Lee C, Ackenhausen A, Spiering S, Fanghanel J . How do spinal segments move?. J Biomech. 2009; 42(14):2286-93. DOI: 10.1016/j.jbiomech.2009.06.055. View

2.
Kumar N, Kukreti S, Ishaque M, Sengupta D, Mulholland R . Functional anatomy of the deer spine: an appropriate biomechanical model for the human spine?. Anat Rec. 2002; 266(2):108-17. DOI: 10.1002/ar.10041. View

3.
Perez-Orribo L, Zucherman J, Hsu K, Reyes P, Rodriguez-Martinez N, Crawford N . Biomechanics of a Posterior Lumbar Motion Stabilizing Device: In Vitro Comparison to Intact and Fused Conditions. Spine (Phila Pa 1976). 2015; 41(2):E55-63. DOI: 10.1097/BRS.0000000000001148. View

4.
Ahmadi A, Maroufi N, Behtash H, Zekavat H, Parnianpour M . Kinematic analysis of dynamic lumbar motion in patients with lumbar segmental instability using digital videofluoroscopy. Eur Spine J. 2009; 18(11):1677-85. PMC: 2899390. DOI: 10.1007/s00586-009-1147-x. View

5.
Anderst W, Baillargeon E, Donaldson W, Lee J, Kang J . Motion path of the instant center of rotation in the cervical spine during in vivo dynamic flexion-extension: implications for artificial disc design and evaluation of motion quality after arthrodesis. Spine (Phila Pa 1976). 2013; 38(10):E594-601. PMC: 3656913. DOI: 10.1097/BRS.0b013e31828ca5c7. View