A biosensor for 3-hydroxybutyrate (3-HB) involving immobilization of the enzyme 3-hydroxybutyrate dehydrogenase onto a screen-printed carbon electrode modified with reduced graphene oxide (GO) and thionine (THI) is reported here. After addition of 3-hydroxybutyrate or the sample in the presence of NAD⁺ cofactor, the generated NADH could be detected amperometrically at 0.0 V vs. Ag pseudo reference electrode. Under the optimized experimental conditions, a calibration plot for 3-HB was constructed showing a wide linear range between 0.010 and 0.400 mM 3-HB which covers the clinically relevant levels for diluted serum samples. In addition, a limit of detection of 1.0 µM, much lower than that reported using other biosensors, was achieved. The analytical usefulness of the developed biosensor was demonstrated via application to spiked serum samples.

Citing Articles

Functionalized Graphene-Based Biosensors for Early Detection of Subclinical Ketosis in Dairy Cows.

Chick S, Ataei Kachouei M, Knowlton K, Ali M ACS Appl Mater Interfaces. 2024; 16(39):51932-51943.

PMID: 39172629 PMC: 11450706. DOI: 10.1021/acsami.4c07715.


A Methodical Review on the Applications and Potentialities of Using Nanobiosensors for Disease Diagnosis.

Ukhurebor K, Onyancha R, Aigbe U, Uk-Eghonghon G, Kerry R, Kusuma H Biomed Res Int. 2022; 2022:1682502.

PMID: 35103234 PMC: 8799955. DOI: 10.1155/2022/1682502.


Highly Sensitive Electrochemical Sensor for Diagnosis of Diabetic Ketoacidosis (DKA) by Measuring Ketone Bodies in Urine.

Go A, Park S, Ku Y, Sun M, Yeon S, Lee J Sensors (Basel). 2021; 21(14).

PMID: 34300644 PMC: 8309864. DOI: 10.3390/s21144902.


Electrochemical Sensors Based on Carbon Nanomaterial Used in Diagnosing Metabolic Disease.

Zhang C, Du X Front Chem. 2020; 8:651.

PMID: 32850664 PMC: 7432198. DOI: 10.3389/fchem.2020.00651.


Development of an Electrochemical Immunosensor for Specific Detection of Visceral Leishmaniasis Using Gold-Modified Screen-Printed Carbon Electrodes.

Martins B, Barbosa Y, Andrade C, Pereira L, Simao G, de Oliveira C Biosensors (Basel). 2020; 10(8).

PMID: 32717832 PMC: 7460044. DOI: 10.3390/bios10080081.


References
1.
Kwan R, Hon P, Mak W, Law L, Hu J, Renneberg R . Biosensor for rapid determination of 3-hydroxybutyrate using bi-enzyme system. Biosens Bioelectron. 2005; 21(7):1101-6. DOI: 10.1016/j.bios.2005.04.005. View

2.
Zhang J, Yang H, Shen G, Cheng P, Zhang J, Guo S . Reduction of graphene oxide via L-ascorbic acid. Chem Commun (Camb). 2010; 46(7):1112-4. DOI: 10.1039/b917705a. View

3.
Shimomura T, Sumiya T, Ono M, Ito T, Hanaoka T . A novel, disposable, screen-printed amperometric biosensor for ketone 3-β-hydroxybutyrate fabricated using a 3-β-hydroxybutyrate dehydrogenase-mesoporous silica conjugate. Anal Bioanal Chem. 2012; 405(1):297-305. DOI: 10.1007/s00216-012-6494-5. View

4.
Forrow N, Sanghera G, Walters S, Watkin J . Development of a commercial amperometric biosensor electrode for the ketone D-3-hydroxybutyrate. Biosens Bioelectron. 2005; 20(8):1617-25. DOI: 10.1016/j.bios.2004.07.009. View

5.
Chen Y, Zhang X, Yu P, Ma Y . Stable dispersions of graphene and highly conducting graphene films: a new approach to creating colloids of graphene monolayers. Chem Commun (Camb). 2009; (30):4527-9. DOI: 10.1039/b907723e. View