» Articles » PMID: 28988075

Feasibility of a Portable X-ray Fluorescence Device for Bone Lead Measurements of Condor Bones

Overview
Date 2017 Oct 9
PMID 28988075
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

Lead based ammunition is a primary source of lead exposure, especially for scavenging wildlife. Lead poisoning remains the leading cause of diagnosed death for the critically endangered California condors, which are annually monitored via blood tests for lead exposure. The results of these tests are helpful in determining recent exposure in condors and in defining the potential for exposure to other species including humans. Since condors are victim to acute and chronic lead exposure, being able to measure both would lend valuable information on the rates of exposure and accumulation through time. A commercial portable X-ray fluorescence (XRF) device has been optimized to measure bone lead in vivo in humans, but this device could also be valuable for field measurements of bone lead in avian species. In this study, we performed measurements of bone Pb in excised, bare condor bones using inductively coupled plasma mass spectrometry (ICP-MS), a cadmium 109 (Cd-109) K-shell X-ray fluorescence (KXRF) system, and a portable XRF system. Both KXRF and portable XRF bone Pb measurement techniques demonstrated good correlations with ICP-MS results (r=0.93 and r=0.92 respectively), even with increasing skin thickness (r=0.86 between ICP-MS and portable XRF at 1.54mm of soft tissue). In conclusion, our results suggest that a portable XRF could be a useful option for measurement of bone Pb in avian species in the field.

Citing Articles

Ways to Measure Metals: From ICP-MS to XRF.

Adesina K, Burgos C, Grier T, Sayam A, Specht A Curr Environ Health Rep. 2025; 12(1):7.

PMID: 39865194 DOI: 10.1007/s40572-025-00473-y.


Advancing the application of pXRF for animal samples.

Brandis K, Francis R, Zawada K, Hasselerharm C, Ramp D PLoS One. 2024; 19(12):e0297830.

PMID: 39636899 PMC: 11620409. DOI: 10.1371/journal.pone.0297830.


Bone lead measurements of live condors in field to assess cumulative lead exposure.

Specht A, Sayam A, Parish C, Hauck T, Watson R, McClure C Environ Sci Pollut Res Int. 2024; 31(44):56308-56313.

PMID: 39264498 PMC: 11900942. DOI: 10.1007/s11356-024-34950-y.


Lead exposure across the life course and age at death.

Farmer J, Specht A, Punshon T, Jackson B, Bidlack F, Bakalar C Sci Total Environ. 2024; 927:171975.

PMID: 38547974 PMC: 11069331. DOI: 10.1016/j.scitotenv.2024.171975.


Benchtop x-ray fluorescence to quantify elemental content in nails non-destructively.

Specht A, Adesina K, Read D, Weisskopf M Sci Total Environ. 2024; 918:170601.

PMID: 38309346 PMC: 10923075. DOI: 10.1016/j.scitotenv.2024.170601.


References
1.
Barry P . A comparison of concentrations of lead in human tissues. Br J Ind Med. 1975; 32(2):119-39. PMC: 1008038. DOI: 10.1136/oem.32.2.119. View

2.
Specht A, Lin Y, Weisskopf M, Yan C, Hu H, Xu J . XRF-measured bone lead (Pb) as a biomarker for Pb exposure and toxicity among children diagnosed with Pb poisoning. Biomarkers. 2016; 21(4):347-52. PMC: 4964874. DOI: 10.3109/1354750X.2016.1139183. View

3.
Somervaille L, Nilsson U, Chettle D, Tell I, Scott M, Schutz A . In vivo measurements of bone lead--a comparison of two x-ray fluorescence techniques used at three different bone sites. Phys Med Biol. 1989; 34(12):1833-45. DOI: 10.1088/0031-9155/34/12/007. View

4.
Kollander B, Widemo F, Agren E, Larsen E, Loeschner K . Detection of lead nanoparticles in game meat by single particle ICP-MS following use of lead-containing bullets. Anal Bioanal Chem. 2016; 409(7):1877-1885. DOI: 10.1007/s00216-016-0132-6. View

5.
Church M, Gwiazda R, Risebrough R, Sorenson K, Chamberlain C, Farry S . Ammunition is the principal source of lead accumulated by California condors re-introduced to the wild. Environ Sci Technol. 2006; 40(19):6143-50. DOI: 10.1021/es060765s. View