Prospective Quality Control Study of a Novel Gravity-driven Whole Blood Separation System Suitable for Humanitarian Crises

Overview
Journal Vox Sang
Specialty Hematology
Date 2017 Sep 28
PMID 28952151
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Centrifugation-based whole blood (WB) separation represents the worldwide standard but it depends on electricity and infrastructure. We have prospectively evaluated a novel hollow-fibre WB separation system that does not require manual priming or blood flow regulation (n = 29). RBC units contained sufficient Hb (50·4 g ± 4·3), low leucocytes (90 000 ± 0·008), exhibited low haemolysis (0·57% ± 0·49) and robust ATP content (51·47% ± 8·2) after 43 days storage. Plasma units contained low leucocytes and mean coagulation factor activities for FV, FVIII and FXI were 47%, 90% and 68%, respectively. RBC met quality specifications but plasma units exhibited reduced FV and FXI activity.

Citing Articles

Migration Study of Dielectrophoretically Manipulated Red Blood Cells in Tapered Aluminium Microelectrode Array: A Pilot Study.

Samad M, Ponnuthurai D, Badrudin S, Ali M, Razak M, Buyong M Micromachines (Basel). 2023; 14(8).

PMID: 37630162 PMC: 10457829. DOI: 10.3390/mi14081625.


Prospective Comparative Quality Control Study of a Novel Gravity-Driven Hollow-Fiber Whole Blood Separation System for the Production of Canine Blood Products.

Lehmann H, Hindricks E, Hassdenteufel E, Moritz A, Bauer N Front Vet Sci. 2019; 6:149.

PMID: 31157246 PMC: 6533853. DOI: 10.3389/fvets.2019.00149.


Traditional and emerging technologies for washing and volume reducing blood products.

Lu M, Lezzar D, Voros E, Shevkoplyas S J Blood Med. 2019; 10:37-46.

PMID: 30655711 PMC: 6322496. DOI: 10.2147/JBM.S166316.