Reliability and Validity of Measuring Respiration Movement Using a Wearable Strain Sensor in Healthy Subjects

Overview
Journal J Phys Ther Sci
Date 2017 Sep 22
PMID 28931984
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

[Purpose] The purpose of this research was to verify the reliability and validity of measuring respiration movement using a wearable strain sensor (WSS) which has been developed newly for clinical objective assessment. [Subjects and Methods] The 21 healthy male students were advised to conduct a breathing movement using measuring tape (MT) and WSS respectively, which was the measured at four locations on chest and abdomen wall. The different degree of chest expansion from maximal end of expiration to maximal end of inspiration was confirmed at each location. The intra-rater ICC (1,1) with 95% confidence interval was used to assess the test-retest reliability, Pearson's correlation analysis was performed to establish the validity. [Results] All ICC values for intra-rater reliability were from 0.94 to 0.98 at all locations, which means that there is a high correlation. All values for validity showed significantly positive, indicating that there is a correlation between the measuring tape and WSS at four locations. [Conclusion] Compared to the measuring tape, WSS has been proved to have high reliability and validity. The finding of this research indicated that WSS is reliable to use for objective measurement of respiratory movements on the chest and abdomen wall in clinical assessment.

Citing Articles

Wearable Sensors for Remote Health Monitoring: Potential Applications for Early Diagnosis of Covid-19.

Mirjalali S, Peng S, Fang Z, Wang C, Wu S Adv Mater Technol. 2021; 7(1):2100545.

PMID: 34901382 PMC: 8646515. DOI: 10.1002/admt.202100545.


Advancements in Methods and Camera-Based Sensors for the Quantification of Respiration.

Rehouma H, Noumeir R, Essouri S, Jouvet P Sensors (Basel). 2020; 20(24).

PMID: 33348827 PMC: 7766256. DOI: 10.3390/s20247252.


Clinical evaluation of stretchable and wearable inkjet-printed strain gauge sensor for respiratory rate monitoring at different measurements locations.

Al-Halhouli A, Al-Ghussain L, El Bouri S, Liu H, Zheng D J Clin Monit Comput. 2020; 35(3):453-462.

PMID: 32088910 DOI: 10.1007/s10877-020-00481-3.


Wearable Sensor-Based Digital Biomarker to Estimate Chest Expansion During Sit-to-Stand Transitions-A Practical Tool to Improve Sternal Precautions in Patients Undergoing Median Sternotomy.

Wang C, Goel R, Noun M, Ghanta R, Najafi B IEEE Trans Neural Syst Rehabil Eng. 2019; 28(1):165-173.

PMID: 31714229 PMC: 7027954. DOI: 10.1109/TNSRE.2019.2952076.


Optimization of Geometry Parameters of Inkjet-Printed Silver Nanoparticle Traces on PDMS Substrates Using Response Surface Methodology.

Abu-Khalaf J, Al-Ghussain L, Nadi A, Saraireh R, Rabayah A, Altarazi S Materials (Basel). 2019; 12(20).

PMID: 31614795 PMC: 6829261. DOI: 10.3390/ma12203329.


References
1.
Ragnarsdottir M, Kristinsdottir E . Breathing movements and breathing patterns among healthy men and women 20-69 years of age. Reference values. Respiration. 2005; 73(1):48-54. DOI: 10.1159/000087456. View

2.
Gilbert R, AUCHINCLOSS Jr J, Peppi D . Relationship of rib cage and abdomen motion to diaphragm function during quiet breathing. Chest. 1981; 80(5):607-12. DOI: 10.1378/chest.80.5.607. View

3.
Wang C, Li X, Gao E, Jian M, Xia K, Wang Q . Carbonized Silk Fabric for Ultrastretchable, Highly Sensitive, and Wearable Strain Sensors. Adv Mater. 2016; 28(31):6640-8. DOI: 10.1002/adma.201601572. View

4.
Binazzi B, Lanini B, Bianchi R, Romagnoli I, Nerini M, Gigliotti F . Breathing pattern and kinematics in normal subjects during speech, singing and loud whispering. Acta Physiol (Oxf). 2006; 186(3):233-46. DOI: 10.1111/j.1748-1716.2006.01529.x. View

5.
Aliverti A, Dellaca R, Pelosi P, Chiumello D, Pedotti A, Gattinoni L . Optoelectronic plethysmography in intensive care patients. Am J Respir Crit Care Med. 2000; 161(5):1546-52. DOI: 10.1164/ajrccm.161.5.9903024. View