» Articles » PMID: 28817381

Advanced Radiation DOSimetry Phantom (ARDOS): a Versatile Breathing Phantom for 4D Radiation Therapy and Medical Imaging

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2017 Aug 18
PMID 28817381
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

A novel breathing phantom was designed for being used in conventional and ion-beam radiotherapy as well as for medical imaging. Accurate dose delivery and patient safety are aimed to be verified for four-dimensional (4D) treatment techniques compensating for breathing-induced tumor motion. The phantom includes anthropomorphic components representing an average human thorax. It consists of real tissue equivalent materials to fulfill the requirements for dosimetric experiments and imaging purposes. The different parts of the torso (lungs, chest wall, and ribs) and the tumor can move independently. Simple regular movements, as well as more advanced patient-specific breathing cycles are feasible while a reproducible setup can be guaranteed. The phantom provides the flexibility to use different types of dosimetric devices and was designed in a way that it is robust, transportable and easy to handle. Tolerance levels and the reliability of the phantom setup were determined in combination with tests on motion accuracy and reproducibility by using infrared optical tracking technology. Different imaging was performed including positron emission tomography imaging, 4D computed tomography as well as real-time in-room imaging. The initial dosimetric benchmarking studies were performed in a photon beam where dose parameters are predictable and the dosimetric procedures well established.

Citing Articles

Regression fitting megavoltage depth dose curves to determine material relative electron density in radiotherapy.

Karl A, Steel J, Warr G Phys Eng Sci Med. 2023; 46(4):1387-1397.

PMID: 37733264 DOI: 10.1007/s13246-023-01306-8.


Patient Breathing Motion and Delivery Specifics Influencing the Robustness of a Proton Pancreas Irradiation.

Knausl B, Lebbink F, Fossati P, Engwall E, Georg D, Stock M Cancers (Basel). 2023; 15(9).

PMID: 37174016 PMC: 10177445. DOI: 10.3390/cancers15092550.


A novel bone suppression algorithm in intensity-based 2D/3D image registration for real-time tumor motion monitoring: Development and phantom-based validation.

Gulyas I, Trnkova P, Knausl B, Widder J, Georg D, Renner A Med Phys. 2022; 49(8):5182-5194.

PMID: 35598307 PMC: 9540269. DOI: 10.1002/mp.15716.


The Influence of Motion on the Delivery Accuracy When Comparing Actively Scanned Carbon Ions versus Protons at a Synchrotron-Based Radiotherapy Facility.

Lebbink F, Stock M, Georg D, Knausl B Cancers (Basel). 2022; 14(7).

PMID: 35406558 PMC: 8997550. DOI: 10.3390/cancers14071788.


Additively Manufactured Patient-Specific Anthropomorphic Thorax Phantom With Realistic Radiation Attenuation Properties.

Hatamikia S, Oberoi G, Unger E, Kronreif G, Kettenbach J, Buschmann M Front Bioeng Biotechnol. 2020; 8:385.

PMID: 32457883 PMC: 7225309. DOI: 10.3389/fbioe.2020.00385.