» Articles » PMID: 28663642

Capnographic Analysis of Minimum Mandatory Flow Rate for Hudson Face Mask: A Randomized Double-blind Study

Overview
Specialty Anesthesiology
Date 2017 Jul 1
PMID 28663642
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Oxygen supplementation through Hudson type face mask is frequently used in perioperative settings. Hudson mask is a variable rate performance device with the risk of rebreathing. Studies using capnography to find out an actual fraction of rebreathing in spontaneously breathing patients are not available.

Aims: In this study, we analyzed the effects of different flow rates through Hudson mask with capnography on fractional inspired carbon dioxide (FiCO).

Setting And Designs: Forty patients posted for minor surgeries under monitored anesthesia care were divided into four groups. They received oxygen flow rate of 3 L/min, 4 L/min, 5 L/min, and 6 L/min as per group division, through Hudson mask.

Materials And Methods: Parameters such as pulse rate, noninvasive blood pressure, oxygen saturation, respiratory rate (RR), end-tidal carbon dioxide (EtCO), and FiCO were noted at baseline. After starting oxygen supplementation, these parameters were monitored every minute for ten minutes. Statistical analysis was done using analysis of variance and Kruskal-Wallis test. Pearson correlation was used to establish a relation between RR and FiCO.

Results And Conclusions: EtCO and FiCO were comparable in all four groups with no statistical significance. There was strong positive correlation seen between RR and FiCO at a flow rate of 3 L/min while negative correlation in other groups. We conclude that FiCO is similar at all flow rates while breathing through Hudson mask. Increase in RR can cause increase in FiCO at lower flow rate, which is within physiological tolerance limits.

Citing Articles

Inhaled CO Concentration While Wearing Face Masks: A Pilot Study Using Capnography.

Acuti Martellucci C, Flacco M, Martellucci M, Violante F, Manzoli L Environ Health Insights. 2022; 16:11786302221123573.

PMID: 36133777 PMC: 9483963. DOI: 10.1177/11786302221123573.


Investigating the Effects of Protective Face Masks on the Respiratory Parameters of Children in the Postanesthesia Care Unit During the COVID-19 Pandemic.

Dost B, Komurcu O, Bilgin S, Dokmeci H, Terzi O, Baris S J Perianesth Nurs. 2021; 37(1):94-99.

PMID: 34824000 PMC: 7877201. DOI: 10.1016/j.jopan.2021.02.004.

References
1.
Williams A, Jones P, Mapleson W . A comparison of oxygen therapy devices used in the postoperative recovery period. Anaesthesia. 1988; 43(2):131-5. DOI: 10.1111/j.1365-2044.1988.tb05483.x. View

2.
Campkin N, Ooi R, Soni N . The rebreathing characteristics of the Hudson oxygen mask. Anaesthesia. 1993; 48(3):239-42. DOI: 10.1111/j.1365-2044.1993.tb06911.x. View

3.
Waldau T, Larsen V, Bonde J . Evaluation of five oxygen delivery devices in spontaneously breathing subjects by oxygraphy. Anaesthesia. 1998; 53(3):256-63. DOI: 10.1046/j.1365-2044.1998.00318.x. View

4.
Vercruyssen M, Kamon E, Hancock P . Effects of carbon dioxide inhalation on psychomotor and mental performance during exercise and recovery. Int J Occup Saf Ergon. 2007; 13(1):15-27. DOI: 10.1080/10803548.2007.11076705. View

5.
KORY R, BERGMANN J, SWEET R, Smith J . Comparative evaluation of oxygen therapy techniques. JAMA. 1962; 179:767-72. DOI: 10.1001/jama.1962.03050100021005. View