Bioelectronic Block of Paravertebral Sympathetic Nerves Mitigates Post-myocardial Infarction Ventricular Arrhythmias
Overview
Affiliations
Background: Autonomic dysfunction contributes to induction of ventricular tachyarrhythmia (VT).
Objective: To determine the efficacy of charge-balanced direct current (CBDC), applied to the T1-T2 segment of the paravertebral sympathetic chain, on VT inducibility post-myocardial infarction (MI).
Methods: In a porcine model, CBDC was applied in acute animals (n = 7) to optimize stimulation parameters for sympathetic blockade and in chronic MI animals (n = 7) to evaluate the potential for VTs. Chronic MI was induced by microsphere embolization of the left anterior descending coronary artery. At termination, in anesthetized animals and following thoracotomy, an epicardial sock array was placed over both ventricles and a quadripolar carousel electrode positioned underlying the right T1-T2 paravertebral chain. In acute animals, the efficacy of CBDC carousel (CBDCC) block was assessed by evaluating cardiac function during T2 paravertebral ganglion stimulation with and without CBDCC. In chronic MI animals, VT inducibility was assessed by extrasystolic (S1-S2) stimulations at baseline and under >66% CBDCC blockade of T2-evoked sympathoexcitation.
Results: CBDCC demonstrated a current-dependent and reversible block without impacting basal cardiac function. VT was induced at baseline in all chronic MI animals. One animal died after baseline induction. Of the 6 remaining animals, only 1 was reinducible with simultaneous CBDCC application (P < .002 from baseline). The ventricular effective refractory period (VERP) was prolonged with CBDCC (323 ± 26 ms) compared to baseline (271 ± 32 ms) (P < .05).
Conclusions: Axonal block of the T1-T2 paravertebral chain with CBDCC reduced VT in a chronic MI model. CBDCC prolonged VERP, without altering baseline cardiac function, resulting in improved electrical stability.
Mechanism of Ventricular Tachycardia Occurring in Chronic Myocardial Infarction Scar.
Donahue J, Chrispin J, Ajijola O Circ Res. 2024; 134(3):328-342.
PMID: 38300981 PMC: 10836816. DOI: 10.1161/CIRCRESAHA.123.321553.
Sympathetic Modulation in Cardiac Arrhythmias: Where We Stand and Where We Go.
Chung W, Lin Y, Wu M, Chang K J Pers Med. 2023; 13(5).
PMID: 37240956 PMC: 10221179. DOI: 10.3390/jpm13050786.
Elia A, Fossati S Front Physiol. 2023; 14:1060666.
PMID: 36798942 PMC: 9926972. DOI: 10.3389/fphys.2023.1060666.
Fuzzy Logic Control of Heartrate by Electrical Block of Vagus Nerve.
Bender S, Green D, Daniels R, Kilgore K, Bhadra N, Vrabec T Int IEEE EMBS Conf Neural Eng. 2021; 2021:1083-1086.
PMID: 34909125 PMC: 8667196. DOI: 10.1109/ner49283.2021.9441092.
Scalable and reversible axonal neuromodulation of the sympathetic chain for cardiac control.
Hadaya J, Buckley U, Gurel N, Chan C, Swid M, Bhadra N Am J Physiol Heart Circ Physiol. 2021; 322(1):H105-H115.
PMID: 34860595 PMC: 8714250. DOI: 10.1152/ajpheart.00568.2021.