Difference in Haemodynamic Microenvironment in Vessels Scaffolded with Absorb BVS and Mirage BRMS: Insights from a Preclinical Endothelial Shear Stress Study

Overview
Date 2017 Jun 8
PMID 28590249
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: In silico studies have provided robust evidence that stent design affects local haemodynamic forces, which appear as a major determinant of clinical outcomes following stent implantation. However, the implications of different stent/scaffold configurations on local haemodynamic forces have not yet been investigated in vivo in a comparative fashion. The aim of this study was to compare the ESS distribution in two differently shaped scaffolds using OCT-based modelling.

Methods And Results: Eight healthy mini pigs were implanted with six Absorb everolimus-eluting bioresorbable vascular scaffolds (Absorb BVS) and five Mirage sirolimus-eluting bioresorbable microfibre scaffolds (Mirage BRMS). Optical coherence tomography (OCT) was performed and strut protrusion was assessed post scaffold implantation. Following the reconstruction of coronary anatomy, blood flow simulation was performed and endothelial shear stress (ESS) was estimated on top of the struts and at luminal surface between the struts in each scaffold. The thicker struts in Absorb (152±140 μm) resulted in an increased protruded distance compared to Mirage (117±123 μm) (p=0.003). This had an effect on the local haemodynamic microenvironment. ESS at the top of the struts was higher in Absorb (1.69±1.20 Pa) than in Mirage (1.53±0.91 Pa) (p<0.001), but lower at inter-strut zones (0.60±0.51 Pa vs. 0.63±0.50 Pa; p<0.01) compared to Mirage. Both scaffold types revealed comparable percentages of vessel luminal surface exposed to recirculation.

Conclusions: Absorb demonstrated higher shear stress on top of the struts compared to Mirage. However, in the inter-strut zones shear stress was higher in Mirage than in Absorb. Further research is required to examine the potential value of in vivo computational modelling in optimising scaffold configuration and clinical outcomes.

Citing Articles

Safety and efficacy of a novel 3D-printed bioresorbable sirolimus-eluting scaffold in a porcine model.

Shi Q, Zhang B, Wang X, Fei J, Qin Q, Zheng B AsiaIntervention. 2023; 9(2):133-142.

PMID: 37736208 PMC: 10507451. DOI: 10.4244/AIJ-D-22-00051.


Evaluating the Arteriotomy Size of a New Sutureless Coronary Anastomosis Using a Finite Volume Approach.

Crielaard H, Hoogewerf M, van Putte B, van de Vosse F, Vlachojannis G, Stecher D J Cardiovasc Transl Res. 2023; 16(4):916-926.

PMID: 36943615 PMC: 10480236. DOI: 10.1007/s12265-023-10367-9.


Clinical outcomes with unselected use of an ultrathin-strut sirolimus-eluting stent: a report from the Swedish Coronary Angiography and Angioplasty Registry (SCAAR).

Buccheri S, Sarno G, Erlinge D, Renlund H, Lagerqvist B, Grimfjard P EuroIntervention. 2020; 16(17):1413-1421.

PMID: 33016880 PMC: 9724977. DOI: 10.4244/EIJ-D-20-00429.


Scaffold underexpansion and late lumen loss after bioresorbable scaffold implantation: Insights from ABSORB JAPAN trial.

Okada K, Honda Y, Kitahara H, Ikutomi M, Kameda R, Hollak M Int J Cardiol Heart Vasc. 2020; 31:100623.

PMID: 32944608 PMC: 7481138. DOI: 10.1016/j.ijcha.2020.100623.


The Evolution of Data Fusion Methodologies Developed to Reconstruct Coronary Artery Geometry From Intravascular Imaging and Coronary Angiography Data: A Comprehensive Review.

Kilic Y, Safi H, Bajaj R, Serruys P, Kitslaar P, Ramasamy A Front Cardiovasc Med. 2020; 7:33.

PMID: 32296713 PMC: 7136420. DOI: 10.3389/fcvm.2020.00033.