Purpose: Dual mobility THA implantations are increasing. THA survival of more than 15 years is considered long-term survival based on the Charnley prosthesis, which has a median wear rate of 39 to 98 mm/year. What is the median wear rate of dual mobility THA cups and its repartition on the liner volume? What are the factors influencing dual mobility wear?

Methods: Thirty-five explants were analysed. The liners were entirely mapped with a fully automatic stripe light scanner that uses green LED light.

Results: The liners had a median wear rate of 38 mm/year. The distribution of the wear between the two bearings varied greatly. No factor was found influencing wear volumes.

Conclusions: The median annual wear of dual mobility liners is at least as low as the wear of cemented polyethylene liners and lower than equivalent cementless liners. The two articulations of the dual mobility THA do not cause more wear.

Level Of Evidence: Level IV - case series.

Citing Articles

Comparing Stability, Gait, and Functional Score after 40-mm Dual-Mobility Hip Arthroplasty to 36-mm Head Hip Arthroplasty in Elderly Hip Fracture Patients.

Cha Y, Lee S, Bae J, Kang Y, Baek J, Kang J Clin Orthop Surg. 2025; 17(1):62-70.

PMID: 39912063 PMC: 11791493. DOI: 10.4055/cios24148.


Analysis of an explanted dual mobility cup after 21 years, parabolic wear pattern: A case report.

Fessy M, Fessy A, Viste A SICOT J. 2023; 9:3.

PMID: 36661475 PMC: 9878998. DOI: 10.1051/sicotj/2022052.


A review of the clinical and engineering performance of dual-mobility cups for total hip arthroplasty.

Yang S, Xie F, Cui W, Zhang Y, Jin Z Am J Transl Res. 2021; 13(8):9383-9394.

PMID: 34540057 PMC: 8430055.


Risk factors and modes of failure in the modern dual mobility implant. A systematic review and meta-analysis.

Pai F, Ma H, Chou T, Huang T, Huang K, Tsai S BMC Musculoskelet Disord. 2021; 22(1):541.

PMID: 34126990 PMC: 8204435. DOI: 10.1186/s12891-021-04404-4.


Comparison of radiographic changes in rectangular curved short stem with thin versus thick porous coating for cementless total hip arthroplasty: a retrospective study with a propensity score matching.

Munakata Y, Kuramitsu Y, Usui Y, Okazaki K J Orthop Surg Res. 2021; 16(1):247.

PMID: 33849591 PMC: 8042942. DOI: 10.1186/s13018-021-02397-3.


References
1.
Engh Jr C, Claus A, Hopper Jr R, Engh C . Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop Relat Res. 2002; (393):137-46. DOI: 10.1097/00003086-200112000-00016. View

2.
Andrew T, Berridge D, Thomas A, Duke R . Long-term review of ring total hip arthroplasty. Clin Orthop Relat Res. 1985; (201):111-22. View

3.
Caton J, Prudhon J . Over 25 years survival after Charnley's total hip arthroplasty. Int Orthop. 2011; 35(2):185-8. PMC: 3032109. DOI: 10.1007/s00264-010-1197-z. View

4.
Wroblewski B . 15-21-year results of the Charnley low-friction arthroplasty. Clin Orthop Relat Res. 1986; (211):30-5. View

5.
Philippot R, Farizon F, Camilleri J, Boyer B, Derhi G, Bonnan J . Survival of cementless dual mobility socket with a mean 17 years follow-up. Rev Chir Orthop Reparatrice Appar Mot. 2008; 94(8):e23-7. DOI: 10.1016/j.rco.2007.10.013. View