» Articles » PMID: 28448003

Standardized Method for Measuring Collection Efficiency from Wipe-sampling of Trace Explosives

Overview
Journal J Vis Exp
Date 2017 Apr 28
PMID 28448003
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

One of the limiting steps to detecting traces of explosives at screening venues is effective collection of the sample. Wipe-sampling is the most common procedure for collecting traces of explosives, and standardized measurements of collection efficiency are needed to evaluate and optimize sampling protocols. The approach described here is designed to provide this measurement infrastructure, and controls most of the factors known to be relevant to wipe-sampling. Three critical factors (the applied force, travel distance, and travel speed) are controlled using an automated device. Test surfaces are chosen based on similarity to the screening environment, and the wipes can be made from any material considered for use in wipe-sampling. Particle samples of the explosive 1,3,5-trinitroperhydro-1,3,5-triazine (RDX) are applied in a fixed location on the surface using a dry-transfer technique. The particle samples, recently developed to simulate residues made after handling explosives, are produced by inkjet printing of RDX solutions onto polytetrafluoroethylene (PTFE) substrates. Collection efficiency is measured by extracting collected explosive from the wipe, and then related to critical sampling factors and the selection of wipe material and test surface. These measurements are meant to guide the development of sampling protocols at screening venues, where speed and throughput are primary considerations.

Citing Articles

A New Wipe-Sampling Instrument for Measuring the Collection Efficiency of Trace Explosives Residues.

Robinson E, Sisco E, Staymates M, Lawrence J Anal Methods. 2018; 10(2):204-213.

PMID: 29881468 PMC: 5986101. DOI: 10.1039/C7AY02694C.

References
1.
Verkouteren J . Particle characteristics of trace high explosives: RDX and PETN. J Forensic Sci. 2007; 52(2):335-40. DOI: 10.1111/j.1556-4029.2006.00354.x. View

2.
Verkouteren J, Ritchie N, Gillen G . Use of force-sensing array films to improve surface wipe sampling. Environ Sci Process Impacts. 2014; 15(2):373-80. DOI: 10.1039/c2em30644a. View

3.
Ashley K, Braybrooke G, Jahn S, Brisson M, White K . Analytical Performance Criteria. Standardized surface dust sampling methods for metals, with emphasis on beryllium. J Occup Environ Hyg. 2009; 6(12):D97-100. DOI: 10.1080/15459620903022597. View

4.
Cettier J, Bayle M, Beranger R, Billoir E, Nuckols J, Combourieu B . Efficiency of wipe sampling on hard surfaces for pesticides and PCB residues in dust. Sci Total Environ. 2014; 505:11-21. DOI: 10.1016/j.scitotenv.2014.09.086. View

5.
Jain S, Heiser A, Venter A . Spray desorption collection: an alternative to swabbing for pharmaceutical cleaning validation. Analyst. 2011; 136(7):1298-301. DOI: 10.1039/c0an00728e. View