Assessment of Neuromuscular Transmission by the Evoked Acceleration Response. An Evaluation of the Accuracy of the Acceleration Transducer in Comparison with a Force Displacement Transducer

Overview
Specialty Anesthesiology
Date 1988 Jul 1
PMID 2842996
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents preliminary observations on an acceleration-responsive transducer designed to monitor neuromuscular transmission. Simultaneously evoked acceleration and tension responses of the adductor pollicis muscles were studied. Registrations were obtained during recovery from atracurium-induced block in 29 individuals in neurolept II anaesthesia (Group I) and in 4 ICU patients (Group II) sedated with pentobarbital or midazolam. Regression analysis of 1567 train-of-four (TOF) registrations, in regard to TOF-ratio (T4/T1) and first twitch ratio (T1/T0), demonstrated regression coefficients (b) and correlation coefficients (r) in the range 0.91-1.06 and 0.89-0.98, respectively. During 1 Hz single twitch stimulation and post-tetanic count stimulation, b and r were in the range 0.85-1.03 and 0.77-0.90, respectively. Following administration of edrophonium (n = 6, Group I) a deviation of T1/T0 regression values was observed in four individuals, i.e. 0.48 (b) and 0.56 (r). This investigation proved a good level of accuracy of the acceleration transducer compared to the force displacement transducer during spontaneous and neostigmine-induced recovery from atracurium block. The acceleration transducer-based system does not require a rigid suspension and seems to have a good monitoring potential in clinical assessment of neuromuscular transmission.

Citing Articles

Differential effects of vecuronium on the thumb and the big toe muscles evaluated by acceleration measurement.

Kitajima T, Ishii K, Kobayashi T, Ogata H J Anesth. 2017; 8(2):143-145.

PMID: 28921132 DOI: 10.1007/BF02514701.


Effects of age on neuromuscular blockade by vecuronium as measured by accelography under sevoflurane anesthesia.

Kitajima T, Ishii K, Ogata H J Anesth. 2013; 10(1):88-91.

PMID: 23839563 DOI: 10.1007/BF02482079.


Postoperative residual block in postanesthesia care unit more than two hours after the administration of a single intubating dose of atracurium.

Rahimi Varposhti M, Heidari S, Safavi M, Honarmand A, Raeesi S J Res Med Sci. 2011; 16(5):651-7.

PMID: 22091288 PMC: 3214377.


Posttetanic count revisited: are measurements more reliable using the TOF-Watch accelerographic peripheral nerve stimulator?.

Vincent Jr R, Brockwell R, Moreno M, Adkins S J Clin Monit Comput. 2004; 18(1):33-7.

PMID: 15139581 DOI: 10.1023/b:jocm.0000025277.12455.66.


New equipment for neuromuscular transmission monitoring: a comparison of the TOF-Guard with the Myograph 2000.

Helbo-Hansen H, Lindholm P, Pedersen H, Severinsen I, Schmidt M J Clin Monit Comput. 1998; 14(1):19-27.

PMID: 9641852 DOI: 10.1023/b:jocm.0000012481.06502.02.