Evaluation of Stability of Stereotactic Space Defined by Cone-beam CT for the Leksell Gamma Knife Icon

Overview
Date 2017 Apr 19
PMID 28419781
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

The Gamma Knife Icon comes with an integrated cone-beam CT (CBCT) for image-guided stereotactic treatment deliveries. The CBCT can be used for defining the Leksell stereotactic space using imaging without the need for the traditional invasive frame system, and this allows also for frameless thermoplastic mask stereotactic treatments (single or fractionated) with the Gamma Knife unit. In this study, we used an in-house built marker tool to evaluate the stability of the CBCT-based stereotactic space and its agreement with the standard frame-based stereotactic space. We imaged the tool with a CT indicator box using our CT-simulator at the beginning, middle, and end of the study period (6 weeks) for determining the frame-based stereotactic space. The tool was also scanned with the Icon's CBCT on a daily basis throughout the study period, and the CBCT images were used for determining the CBCT-based stereotactic space. The coordinates of each marker were determined in each CT and CBCT scan using the Leksell GammaPlan treatment planning software. The magnitudes of vector difference between the means of each marker in frame-based and CBCT-based stereotactic space ranged from 0.21 to 0.33 mm, indicating good agreement of CBCT-based and frame-based stereotactic space definition. Scanning 4-month later showed good prolonged stability of the CBCT-based stereotactic space definition.

Citing Articles

Not one size fits all: a half-mask-based GKRS treatment for preservation of a scalp flap. Illustrative case.

Goulenko V, Madhugiri V, Almeida N, Podgorsak M, Fenstermaker R, Prasad D J Neurosurg Case Lessons. 2025; 9(10).

PMID: 40064013 PMC: 11894276. DOI: 10.3171/CASE24852.


Zero Setup Margin Mask versus Frame Immobilization during Gamma Knife Icon™ Stereotactic Radiosurgery for Brain Metastases.

Kutuk T, Kotecha R, Tolakanahalli R, Wieczorek D, Lee Y, Ahluwalia M Cancers (Basel). 2022; 14(14).

PMID: 35884453 PMC: 9320023. DOI: 10.3390/cancers14143392.


Discrepancies between frame- and CBCT-based stereotactic space definition on the Gamma Knife Icon.

Bannikova I, Dalechina A, Kostjuchenko V, ZHuravleva A, Golanov A, Banov S J Appl Clin Med Phys. 2022; 23(7):e13637.

PMID: 35635549 PMC: 9278688. DOI: 10.1002/acm2.13637.


Ramifications of Setup Margin Use During Frameless Stereotactic Radiosurgery/Therapy With Gamma Knife Icon Cone-Beam Computed Tomography (CBCT): A Dosimetric Study.

Duggar W, Morris B, He R, Yang C Cureus. 2022; 14(2):e21996.

PMID: 35282559 PMC: 8906882. DOI: 10.7759/cureus.21996.


Total workflow uncertainty of frameless radiosurgery with the Gamma Knife Icon cone-beam computed tomography.

Duggar W, Morris B, He R, Yang C J Appl Clin Med Phys. 2022; 23(5):e13564.

PMID: 35157361 PMC: 9121051. DOI: 10.1002/acm2.13564.


References
1.
LEKSELL L . The stereotaxic method and radiosurgery of the brain. Acta Chir Scand. 1951; 102(4):316-9. View

2.
Cernica G, Wang Z, Malhotra H, de Boer S, Podgorsak M . Investigation of gamma knife image registration errors resulting from misalignment between the patient and the imaging axis. Med Phys. 2006; 33(4):941-3. DOI: 10.1118/1.2179751. View

3.
Li W, Bootsma G, Von Schultz O, Carlsson P, Laperriere N, Millar B . Preliminary Evaluation of a Novel Thermoplastic Mask System with Intra-fraction Motion Monitoring for Future Use with Image-Guided Gamma Knife. Cureus. 2016; 8(3):e531. PMC: 4829406. DOI: 10.7759/cureus.531. View

4.
Li W, Cho Y, Ansell S, Laperriere N, Menard C, Millar B . The Use of Cone Beam Computed Tomography for Image Guided Gamma Knife Stereotactic Radiosurgery: Initial Clinical Evaluation. Int J Radiat Oncol Biol Phys. 2016; 96(1):214-20. DOI: 10.1016/j.ijrobp.2016.04.011. View