» Articles » PMID: 28372556

A Randomized Cross-over Study Comparing the Performance of HD Integra™ Central Concentrate System Versus Pre-produced Concentrate in Hemodialysis

Overview
Journal BMC Nephrol
Publisher Biomed Central
Specialty Nephrology
Date 2017 Apr 5
PMID 28372556
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Pre-produced bicarbonate concentrates (PPC) are still widely used in developing countries despite its cost and risk but Central Concentrate System (CCS) is lacking in data to support its wider adoption.

Methods: We conducted an 8-week randomized crossover study on 16 Hemodialysis machines to compare CCS versus PPC. Performance is assessed by solute concentrations while safety is assessed by microbial count, endotoxin level and adverse event reporting.

Results: Microbial counts and endotoxin levels were monitored on 48 occasions during the 8-week study for the CCS arm of the study. The levels were all below the action limit during the study. No patient reported any adverse events. Dialysate Sodium, Chloride and Bicarbonate concentrations were measured on a total of 128 occasions for each arm of the study. The relative deviations of Sodium, Chloride and Bicarbonate concentration were within ±5% of their nominal values for both. The 95% Confidence Intervals for the ratio of the mean solute concentrations on the CCS to PPC lie within the tolerance limit of ±5%.

Conclusion: Modern CCS is bacteriologically safe and its performance statistically equivalent to PPC.

References
1.
Stragier A, Wenderickx D . Bacterial growth prevention in liquid bicarbonate concentrate. EDTNA ERCA J. 1999; 24(3):40-2, 44. View

2.
De Vos J, Marzougui H, Hombrouckx R . Automated on-line bicarbonate concentrate production: microbiological, chemical and economic benefits. EDTNA ERCA J. 1996; 22(4):5-6. View

3.
Rehmert N, Bocker T, Raidt H, Langer K, Graefe H . A new central supply system as alternative source for bicarbonate dialysate. Int J Artif Organs. 1992; 15(6):339-42. View

4.
Ebben J, Hirsch D, Luehmann D, Collins A, Keshaviah P . Microbiologic contamination of liquid bicarbonate concentrate for hemodialysis. ASAIO Trans. 1987; 33(3):269-73. View

5.
Ledebo I . On-line preparation of solutions for dialysis: practical aspects versus safety and regulations. J Am Soc Nephrol. 2002; 13 Suppl 1:S78-83. View