» Articles » PMID: 28366210

3D Printed LED Based On-capillary Detector Housing with Integrated Slit

Overview
Journal Anal Chim Acta
Publisher Elsevier
Specialty Chemistry
Date 2017 Apr 4
PMID 28366210
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

A 3D printed photometric detector body with integrated slit was fabricated to position a LED and photodiode either side of capillary tubing using a fused deposition modelling (FDM) printer. To make this approach suitable for capillaries down to 50 μm i.d. the dimension of the in-built slit is the critical element of the printed housing. The spatial orientation of the model for printing was found to significantly impact on the resolution of the structures and voids that can be printed. By designing a housing with a slit positioned in the XY plane in parallel with the print direction, the narrowest void (slit) that could be printed was 70 μm. The potential use of the 3D printed slit for photometric detection was characterised using tubing and capillary from 500 down to 50 μm i.d, demonstrating a linear response from 632 to 40 mAU. The effective pathlength and stray light varied from 383 to 22 μm and 3.8% - 50% for 500- 50 μm i.d tubing and capillary. The use of a V-shaped alignment feature allowed for easy and reliable positioning of the tubing inside the detector, as demonstrated by a RSD of 1.9% (n = 10) in peak height when repositioning the tubing between measurements using flow injection analysis (FIA). The performance of the 3D printed housing and 70 μm slit was benchmarked against a commercially available interface using the CE separation of Zn and Cu complexes with PAR. The limit of detection with the 3D printed slit was 6.8 and 4.5 μM and is 2.8 and 1.6 μM with the commercial interface.

Citing Articles

Fabrication of Low-Cost Miniaturized Gas Cells via SLA 3D-Printing for UV-Based Gas Sensors.

Barreto D, Gelamo R, Mizaikoff B, da Silveira Petruci J ACS Omega. 2024; 9(7):8374-8380.

PMID: 38405469 PMC: 10883014. DOI: 10.1021/acsomega.3c09317.


Droplet-based μChopper device with a 3D-printed pneumatic valving layer and a simple photometer for absorbance based fructosamine quantification in human serum.

Kayirangwa Y, Mohibullah M, Easley C Analyst. 2023; 148(19):4810-4819.

PMID: 37605899 PMC: 10530610. DOI: 10.1039/d3an01149f.


Past, current, and future roles of 3D printing in the development of capillary electrophoresis systems.

Esene J, Nasman P, Akuoko Y, Tahir A, Woolley A Trends Analyt Chem. 2023; 162.

PMID: 37008739 PMC: 10062378. DOI: 10.1016/j.trac.2023.117032.


Current and future trends of additive manufacturing for chemistry applications: a review.

Alimi O, Meijboom R J Mater Sci. 2021; 56(30):16824-16850.

PMID: 34413542 PMC: 8363067. DOI: 10.1007/s10853-021-06362-7.


[Multimaterial 3D-printed contactless conductivity/laser-induced fluorescence dual-detection cell for capillary electrophoresis].

Zhang P, Yang L, Liu Q, Lu S, Liang Y, Zhang M Se Pu. 2021; 39(8):921-926.

PMID: 34212593 PMC: 9404044. DOI: 10.3724/SP.J.1123.2021.02021.