» Articles » PMID: 28348717

Cephalomedullary Nail Versus Sliding Hip Screw for Fixation of AO 31 A1/2 Intertrochanteric Femoral Fracture: a 12-year Comparison of Failure, Complications, and Mortality

Overview
Specialty Orthopedics
Date 2017 Mar 29
PMID 28348717
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Background: In the United States intertrochanteric and pertrochanteric fractures occur at a rate of more than 150,000 cases annually. Current standard of care for these fractures includes fixation with either a cephalomedullary nail (CMN) or a sliding hip screw (SHS). The purpose of this study was to compare failure and medical complications of intertrochanteric femoral fractures repaired by CMN or SHS.

Methods: This study is a retrospective cohort study that included 249 patients with AO/OTA 31 A1.1-3, 31 A2.1-3 nonpathological fractures of the femur, of which 137 received CMN and 112 received SHS. Analysis was stratified by fracture type as stable (AO 31A1.1-2.1) or unstable (AO 31A2.2-3).

Results: The tip-apex distance in stable fractures fixed with CMN was 17.3 ± 5.9 compared to 26.2 ± 7.9 in the stable SHS group (p<0.001) while it was 19.0 ± 5.3 in the unstable CMN group compared to 24.0 ± 6.7 in the unstable SHS patients ( = 0.004). Among patients with stable fracture patterns there was no difference in collapse, complications, failure, or mortality (all > 0.05). Among patients with unstable fractures CMN had significantly less collapse ( < 0.001) and failure ( = 0.046) but no difference in complications ( = 0.126) or mortality ( = 0.586).

Conclusions: There were no significant differences in failure or complication rates when comparing the CMN to the SHS in stable intertrochanteric fractures. CMN demonstrated significantly reduced failure and collapse rates in unstable intertrochanteric fractures when compared to SHS; however, this study had a relatively small sample size of unstable fractures and all results must be interpreted within this context.

Citing Articles

Exploring the Efficacy of Computer-Assisted Navigation in Improving Lag Screw Placement and Preventing Cut-Out in Intramedullary Nail Fixation of Femoral Fractures: A Meta-Analysis.

Richa P, DeFrancisis J, Young V, Habib F, Danahy P Cureus. 2025; 17(1):e77724.

PMID: 39981447 PMC: 11841961. DOI: 10.7759/cureus.77724.


Single-centre results of a randomised controlled trial comparing the Gamma3 nail and a sliding hip screw to treat AO type 31-A1 and 31-A2 trochanteric fractures.

Hempel E, Wendlandt R, Unger A, Frese J, Wilde E, Schulz A Langenbecks Arch Surg. 2024; 409(1):310.

PMID: 39419825 DOI: 10.1007/s00423-024-03505-7.


Comparison of Outcomes of Long Versus Short Cephalomedullary Nails for the Fixation of Intertrochanteric Femur Fractures: A Systematic Review and Meta-analysis of 14,547 Patients.

Rajnish R, Srivastava A, Kumar P, Yadav S, Sharma S, Haq R Indian J Orthop. 2023; 57(8):1165-1187.

PMID: 37525736 PMC: 10386979. DOI: 10.1007/s43465-023-00915-5.


Morbidity and Mortality Analysis in the Treatment of Intertrochanteric Hip Fracture with Two Fixation Systems: Dynamic Hip Screw (DHS) or Trochanteric Fixation Nail Advance (TFNA).

Lopez-Hualda A, Arruti-Perez E, Bebea-Zamorano F, Sosa-Reina M, Villafane J, Martinez-Martin J Geriatrics (Basel). 2023; 8(3).

PMID: 37367098 PMC: 10298446. DOI: 10.3390/geriatrics8030066.


Case report: unique failure of a Synthes TFNA fenestrated lag screw in a peritrochanteric nonunion.

Smidt K, Perry K, Andre L, Chauvin B, Kautz S, Massey P OTA Int. 2022; 5(2):e203.

PMID: 35919107 PMC: 9278917. DOI: 10.1097/OI9.0000000000000203.


References
1.
Min W, Kim S, Kim T, Lee K, Cho M, Ha Y . Proximal femoral nail for the treatment of reverse obliquity intertrochanteric fractures compared with gamma nail. J Trauma. 2007; 63(5):1054-60. DOI: 10.1097/01.ta.0000240455.06842.a0. View

2.
Burge R, Dawson-Hughes B, Solomon D, Wong J, King A, Tosteson A . Incidence and economic burden of osteoporosis-related fractures in the United States, 2005-2025. J Bone Miner Res. 2006; 22(3):465-75. DOI: 10.1359/jbmr.061113. View

3.
Olson S, Schemitsch G, Morwood M, Schemitsch E, Russell T, Latta L . Hot Topics in Biomechanics: Hip Fracture Fixation. J Orthop Trauma. 2015; 29 Suppl 12:S1-5. DOI: 10.1097/BOT.0000000000000471. View

4.
Aros B, Tosteson A, Gottlieb D, Koval K . Is a sliding hip screw or im nail the preferred implant for intertrochanteric fracture fixation?. Clin Orthop Relat Res. 2008; 466(11):2827-32. PMC: 2565060. DOI: 10.1007/s11999-008-0285-5. View

5.
Park S, Kang J, Kim H, Lee W, Kim Y . Treatment of intertrochanteric fracture with the Gamma AP locking nail or by a compression hip screw--a randomised prospective trial. Int Orthop. 1998; 22(3):157-60. PMC: 3619600. DOI: 10.1007/s002640050231. View