Abstract

Background/aim: In research and development of biliary plastic stents (PS), continuous efforts have been made to overcome short patency time and high rate of migration. The aim of this study was to evaluate the patency and migration rate of different PS shapes for a given period of time.

Methods: Using an in vitro bile phantom model, we compared the patency among different shapes of PS (three straight PS, four double-pigtail PS, and a new screw-shaped PS). We performed an analysis of the degree of luminal narrowing by light microscopic examination. Using an in vivo swine model, we compared the patency and migration rate among the three different types of PS.

Results: Eight weeks after the bile exposure in the bile flow phantom model, 80 PS were retrieved and analyzed. The straight PS showed less biofilm formation and luminal narrowing than other types of PS (p < 0.05). Forty-nine PS were inserted into the dilated bile ducts of 10 swine models, and 39 PS were successfully retrieved 8 weeks later. The stent migration occurred less frequently in the double-pigtail PS and the screw-shaped PS than it did in the straight PS (11.1, 10, and 27.3%, respectively). However, there was no statistical difference in stent patency among the different shapes.

Conclusions: Stent patency may not be significantly different depending on the shape of PS for 8 weeks. The screw-shaped PS showed similar patency and migration rate to the double-pigtail PS. These results may help guiding future PS development and clinical decisions.

Citing Articles

Optimal reproduction of a porcine benign biliary stricture model using endobiliary radiofrequency ablation.

Kwon C, Jang S, Jeong S, Kim G, Lee T, Cho J Sci Rep. 2022; 12(1):12046.

PMID: 35835985 PMC: 9283468. DOI: 10.1038/s41598-022-16340-x.


Crystalline Characteristics and Their Influence in the Mechanical Performance in Poly(ε-Caprolactone) / High Density Polyethylene Blends.

Blazquez-Blazquez E, Perez E, Lorenzo V, Cerrada M Polymers (Basel). 2019; 11(11).

PMID: 31766237 PMC: 6918181. DOI: 10.3390/polym11111874.


Evaluation of valve function in antireflux biliary metal stents.

Kwon C, Moon J, Yun H, Jeong S, Koh D, Lee W BMC Gastroenterol. 2018; 18(1):150.

PMID: 30340463 PMC: 6194677. DOI: 10.1186/s12876-018-0878-8.

References
1.
Katsinelos P, Galanis I, Pilpilidis I, Paroutoglou G, Tsolkas P, Papaziogas B . The effect of indwelling endoprosthesis on stone size or fragmentation after long-term treatment with biliary stenting for large stones. Surg Endosc. 2003; 17(10):1552-5. DOI: 10.1007/s00464-002-9240-9. View

2.
Costamagna G, Mutignani M, Rotondano G, Cipolletta L, Ghezzo L, Foco A . Hydrophilic hydromer-coated polyurethane stents versus uncoated stents in malignant biliary obstruction: a randomized trial. Gastrointest Endosc. 2000; 51(1):8-11. DOI: 10.1016/s0016-5107(00)70378-6. View

3.
Han J, Moon J, Koo H, Kang J, Choi J, Jeong S . Effect of biliary stenting combined with ursodeoxycholic acid and terpene treatment on retained common bile duct stones in elderly patients: a multicenter study. Am J Gastroenterol. 2009; 104(10):2418-21. DOI: 10.1038/ajg.2009.303. View

4.
Terruzzi V, Comin U, De Grazia F, Toti G, Zambelli A, Beretta S . Prospective randomized trial comparing Tannenbaum Teflon and standard polyethylene stents in distal malignant biliary stenosis. Gastrointest Endosc. 2000; 51(1):23-7. DOI: 10.1016/s0016-5107(00)70381-6. View

5.
Choe J, Hyun J . Is There Room for Improvement in Increasing the Stent Patency of Biliary Plastic Stents?. Gut Liver. 2016; 10(4):499-501. PMC: 4933407. DOI: 10.5009/gnl16251. View