Reduction of Metal Artifacts Due to Dental Hardware in Computed Tomography Angiography: Assessment of the Utility of Model-based Iterative Reconstruction

Overview
Journal Neuroradiology
Specialties Neurology
Radiology
Date 2017 Mar 4
PMID 28255903
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The aim of this study is to assess the value of adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR) for reduction of metal artifacts due to dental hardware in carotid CT angiography (CTA).

Methods: Thirty-seven patients with dental hardware who underwent carotid CTA were included. CTA was performed with a GE Discovery CT750 HD scanner and reconstructed with filtered back projection (FBP), ASIR, and MBIR. We measured the standard deviation at the cervical segment of the internal carotid artery that was affected most by dental metal artifacts (SD) and the standard deviation at the common carotid artery that was not affected by the artifact (SD). We calculated the artifact index (AI) as follows: AI = [(SD)2 - (SD)2]1/2 and compared each AI for FBP, ASIR, and MBIR. Visual assessment of the internal carotid artery was also performed by two neuroradiologists using a five-point scale for each axial and reconstructed sagittal image. The inter-observer agreement was analyzed using weighted kappa analysis.

Results: MBIR significantly improved AI compared with FBP and ASIR (p < 0.001, each). We found no significant difference in AI between FBP and ASIR (p = 0.502). The visual score of MBIR was significantly better than those of FBP and ASIR (p < 0.001, each), whereas the scores of ASIR were the same as those of FBP. Kappa values indicated good inter-observer agreements in all reconstructed images (0.747-0.778).

Conclusions: MBIR resulted in a significant reduction in artifact from dental hardware in carotid CTA.

Citing Articles

Evaluation of the effect of reducing metal artifacts in multi-detector CT imaging of zirconia and titanium implants.

Suito H, Yoshihara H, Maeda N, Kasai R, Inoue T, Amano M Oral Radiol. 2025; .

PMID: 40038165 DOI: 10.1007/s11282-025-00814-5.


Image Follow-Up After Flow Diverter Treatment Using Only Ultra-High Resolution CT Angiography with Model-Based Iterative Reconstruction: A Case Report.

Iwata N, Sakamoto M, Sakou T, Uno T, Kurosaki M Yonago Acta Med. 2024; 67(3):254-258.

PMID: 39176194 PMC: 11335917. DOI: 10.33160/yam.2024.08.006.


Reduction of metal artefacts from bilateral hip prostheses during lower extremity computed tomography angiography: an experimental phantom study.

Abdalqader O, Hjouj M, Aljamal M, Hjouj F, Abuzaid M, Mousa M J Med Radiat Sci. 2024; 71(3):421-431.

PMID: 38941235 PMC: 11569399. DOI: 10.1002/jmrs.797.


Utility of follow-up ultra-high-resolution CT angiography with model-based iterative reconstruction after flow diverter treatment for cerebral aneurysms.

Iwata N, Sakamoto M, Sakou T, Uno T, Kurosaki M Radiol Med. 2023; 128(10):1262-1270.

PMID: 37658197 DOI: 10.1007/s11547-023-01692-9.


Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions.

Kawauchi S, Chida K, Hamada Y, Tsuruta W Radiol Phys Technol. 2021; 15(1):25-36.

PMID: 34796447 DOI: 10.1007/s12194-021-00644-0.


References
1.
Borisch I, Boehme T, Butz B, Hamer O, Feuerbach S, Zorger N . Screening for carotid injury in trauma patients: image quality of 16-detector-row computed tomography angiography. Acta Radiol. 2007; 48(7):798-805. DOI: 10.1080/02841850701422104. View

2.
Brook O, Gourtsoyianni S, Brook A, Mahadevan A, Wilcox C, Raptopoulos V . Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology. 2012; 263(3):696-705. DOI: 10.1148/radiol.12111170. View

3.
Huang J, Kerns J, Nute J, Liu X, Balter P, Stingo F . An evaluation of three commercially available metal artifact reduction methods for CT imaging. Phys Med Biol. 2015; 60(3):1047-67. PMC: 4311882. DOI: 10.1088/0031-9155/60/3/1047. View

4.
Morsbach F, Wurnig M, Kunz D, Krauss A, Schmidt B, Kollias S . Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions. Eur Radiol. 2013; 23(10):2687-94. DOI: 10.1007/s00330-013-2885-z. View

5.
Kuya K, Shinohara Y, Sakamoto M, Iwata N, Kishimoto J, Fujii S . CT angiography after carotid artery stenting: assessment of the utility of adaptive statistical iterative reconstruction and model-based iterative reconstruction. Neuroradiology. 2014; 56(11):947-53. DOI: 10.1007/s00234-014-1415-2. View