In Vivo Dosimetry in UK External Beam Radiotherapy: Current and Future Usage

Overview
Journal Br J Radiol
Specialty Radiology
Date 2017 Feb 17
PMID 28205452
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Towards Safer Radiotherapy recommended that radiotherapy (RT) centres should have protocols in place for in vivo dosimetry (IVD) monitoring at the beginning of patient treatment courses (Donaldson S. Towards safer radiotherapy. R Coll Radiol 2008). This report determines IVD implementation in the UK in 2014, the methods used and makes recommendations on future use.

Methods: Evidence from peer-reviewed journals was used in conjunction with the first survey of UK RT centre IVD practice since the publication of Towards Safer Radiotherapy. In March 2014, profession-specific questionnaires were sent to radiographer, clinical oncologist and physics staff groups in each of the 66 UK RT centres.

Results: Response rates from each group were 74%, 45% and 74%, respectively. 73% of RT centres indicated that they performed IVD. Diodes are the most popular IVD device. Thermoluminescent dosimeter (TLD) is still in use in a number of centres but not as a sole modality, being used in conjunction with diodes and/or electronic portal imaging device (EPID). The use of EPID dosimetry is increasing and is considered of most potential value for both geometric and dosimetric verification.

Conclusion: Owing to technological advances, such as electronic data transfer, independent monitor unit checking and daily image-guided radiotherapy, the overall risk of adverse treatment events in RT has been substantially reduced. However, the use of IVD may prevent a serious radiation incident. Point dose IVD is not considered suited to the requirements of verifying advanced RT techniques, leaving EPID dosimetry as the current modality likely to be developed as a future standard. Advances in knowledge: An updated perspective on UK IVD use and provision of professional guidelines for future implementation.

Citing Articles

Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments.

Decker S, Bruza P, Zhang R, Williams B, Jarvis L, Pogue B Med Phys. 2024; 51(8):5754-5763.

PMID: 38598093 PMC: 11321933. DOI: 10.1002/mp.17071.


The role of EPID in vivo dosimetry in the risk management of stereotactic lung treatments.

Esposito M, Mancosu P, Bruschi A, Ghirelli A, Pini S, Alpi P Strahlenther Onkol. 2023; 199(11):992-999.

PMID: 37256302 DOI: 10.1007/s00066-023-02081-x.


Impact through versatility: Patterns of dosimetry utilization with TLD across a large multi-site radiotherapy department.

Lonski P, Kron T Front Oncol. 2022; 12:918178.

PMID: 36330466 PMC: 9623149. DOI: 10.3389/fonc.2022.918178.


Assessment of a Therapeutic X-ray Radiation Dose Measurement System Based on a Flexible Copper Indium Gallium Selenide Solar Cell.

Shin D, Kim T, Rah J, Kim D, Yang H, Lee S Sensors (Basel). 2022; 22(15).

PMID: 35957376 PMC: 9370937. DOI: 10.3390/s22155819.


Comparing treatment uncertainty for ultra- vs. standard-hypofractionated breast radiation therapy based on in-vivo dosimetry.

Fiagan Y, Bossuyt E, Machiels M, Nevens D, Billiet C, Poortmans P Phys Imaging Radiat Oncol. 2022; 22:85-90.

PMID: 35602547 PMC: 9117915. DOI: 10.1016/j.phro.2022.05.003.


References
1.
Wendling M, McDermott L, Mans A, Sonke J, Van Herk M, Mijnheer B . A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments. Med Phys. 2009; 36(7):3310-21. DOI: 10.1118/1.3148482. View

2.
Olaciregui-Ruiz I, Rozendaal R, Mijnheer B, van Herk M, Mans A . Automatic in vivo portal dosimetry of all treatments. Phys Med Biol. 2013; 58(22):8253-64. DOI: 10.1088/0031-9155/58/22/8253. View

3.
Narayanasamy G, Zalman T, Ha C, Papanikolaou N, Stathakis S . Evaluation of Dosimetry Check software for IMRT patient-specific quality assurance. J Appl Clin Med Phys. 2015; 16(3):5427. PMC: 5690116. DOI: 10.1120/jacmp.v16i3.5427. View

4.
Edwards C, Hamer E, Mountford P, Moloney A . An update survey of UK in vivo radiotherapy dosimetry practice. Br J Radiol. 2007; 80(960):1011-4. DOI: 10.1259/bjr/14945156. View

5.
Mackay R, Williams P . The cost effectiveness of in vivo dosimetry is not proven. Br J Radiol. 2009; 82(976):265-6. DOI: 10.1259/bjr/58443203. View