Abstract

Background: The efficacy of devices for difficult intubation in paediatric patients, especially with a Cormack-Lehane grade 4 view, has yet to be established. We compared intubating parameters among three devices (the Airtraq®, McGrath®, and Macintosh laryngoscopes).

Methods: This study is a randomised cross-over trial. Participants were 20 anaesthetists. Each device was tested three times using a paediatric manikin with a Cormack-Lehane grade 4 view. The order to use each device was randomised by a computer-generated random sequence. The primary endpoint was the rate of successful intubation. Secondary endpoints included the time taken to intubate, percentage of glottic opening score, and severity of potential dental trauma.

Results: The successful intubation rates of the Airtraq®, McGrath®, and Macintosh laryngoscopes were 100%, 72%, and 45%, respectively. The risk ratio of the success rates of Airtraq® compared with McGrath® and Macintosh laryngoscopes were 1.40 (95% CI; 1.19-1.64, P < 0.001) and 2.22 (95% CI; 1.68-2.94, P < 0.001), respectively. The modified Cormack-Lehane grade and percentage of the glottic opening score were better for the Airtraq® than for the other devices. The dental trauma score was lower for the Airtraq® than for the other devices. There were no significant differences in the intubation time among the groups.

Conclusions: The Airtraq® had higher success rate, had better visibility, and was associated with less dental trauma than the other devices in a difficult paediatric intubation model with a Cormack-Lehane grade 4 view.

Citing Articles

Clinically Preferred Videolaryngoscopes in Airway Management: An Updated Systematic Review.

Nedunchezhian V, Nedunchezhian I, van Zundert A Healthcare (Basel). 2023; 11(17).

PMID: 37685417 PMC: 10487223. DOI: 10.3390/healthcare11172383.


Comparison of McGrath Videolaryngoscope and Macintosh Laryngoscope in Children with Torticollis: Randomized Controlled Trial.

Hur M, Kim J, Min S, Lee K, Won Y, Kim J Children (Basel). 2021; 8(12).

PMID: 34943365 PMC: 8699989. DOI: 10.3390/children8121171.


Difficult Airway Management in Neonates and Infants: Knowledge of Devices and a Device-Oriented Strategy.

Sawa T, Kainuma A, Akiyama K, Kinoshita M, Shibasaki M Front Pediatr. 2021; 9:654291.

PMID: 34026688 PMC: 8138561. DOI: 10.3389/fped.2021.654291.


The management of unanticipated difficult airways in children of all age groups in anaesthetic practice - the position paper of an expert panel.

Walas W, Aleksandrowicz D, Kornacka M, Gaszynski T, Helwich E, Migdal M Scand J Trauma Resusc Emerg Med. 2019; 27(1):87.

PMID: 31533787 PMC: 6751579. DOI: 10.1186/s13049-019-0666-7.


Comparative evaluation of Airtraq™ and GlideScope® videolaryngoscopes for difficult pediatric intubation in a Pierre Robin manikin.

Desai N, Johnson M, Priddis K, Ray S, Chigaru L Eur J Pediatr. 2019; 178(7):1105-1111.

PMID: 31119438 DOI: 10.1007/s00431-019-03396-7.

References
1.
Maharaj C, Higgins B, Harte B, Laffey J . Evaluation of intubation using the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy--a manikin study. Anaesthesia. 2006; 61(5):469-77. DOI: 10.1111/j.1365-2044.2006.04547.x. View

2.
Maharaj C, Costello J, Higgins B, Harte B, Laffey J . Learning and performance of tracheal intubation by novice personnel: a comparison of the Airtraq and Macintosh laryngoscope. Anaesthesia. 2006; 61(7):671-7. DOI: 10.1111/j.1365-2044.2006.04653.x. View

3.
Vlatten A, Fielding A, Bernard A, Litz S, MacManus B, Soder C . Comparison of the airtraq laryngoscope to the direct laryngoscopy in the pediatric airway. J Pediatr Intensive Care. 2019; 1(2):71-76. PMC: 6530678. DOI: 10.3233/PIC-2012-013. View

4.
Yentis S, Lee D . Evaluation of an improved scoring system for the grading of direct laryngoscopy. Anaesthesia. 1999; 53(11):1041-4. DOI: 10.1046/j.1365-2044.1998.00605.x. View

5.
Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S . Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010; 110(4):1049-55. DOI: 10.1213/ANE.0b013e3181d2aad7. View