Impact of PET Acquisition Durations on Image Quality and Lesion Detectability in Whole-body Ga-PSMA PET-MRI

Overview
Journal EJNMMI Res
Date 2017 Feb 8
PMID 28168589
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Background: While Ga-PSMA PET-MRI might be superior to PET-CT with regard to soft tissue assessment in prostate cancer evaluation, it is also known to potentially introduce additional PET image artefacts. Therefore, the impact of PET acquisition duration and attenuation data on artefact occurrence, lesion detectability, and quantification was investigated. To this end, whole-body PET list mode data from 12 patients with prostate cancer were acquired 1 h after injection of 2 MBq/kg [Ga]HBED-CC-PSMA on a hybrid PET-MRI system. List mode data were further transformed into data sets representing 300, 180, 90, and 30 s acquisition duration per bed position. Standard attenuation and scatter corrections were performed based on MRI-derived attenuation maps, complemented by emission-based attenuation data in areas not covered by MRI. A total of 288 image data sets were reconstructed with varying acquisition durations for emission and attenuation data with and without scatter and prompt gamma correction, and further analysed regarding image quality and diagnostic performance.

Results: Decreased PET acquisition durations resulted in a significantly increased incidence of halo artefacts around kidneys and bladder, decreased lesion detectability and lower SUV as well as markedly lower arm attenuation values: Halo artefacts were present in 5 out of 12 cases at 300-s duration, in 6 at 180 s, in 10 at 90 s, and in 11 cases at 30 s. Using attenuation data of the 300 s scans restored artefact occurrence to the original 300-s level. Prompt gamma correction only led to small improvements in terms of artefact occurrence and size. Of the 141 detected lesions in the 300-s images one lesion was not detected at 180 s, 28 at 90 s, and 64 at 30 s. Using the 300-s attenuation map decreased non-detectability of lesions to zero at 180 s, 9 at 90 s, and 52 at 30 s. Attenuation maps at 90 and 30 s demonstrated markedly lower mean arm attenuation values (0.002 cm) than those at 300 s (0.084 cm), and 180 s (0.062 cm).

Conclusions: Short acquisition durations of less than 3 minutes per bed position result in unacceptable image artefacts and decreased diagnostic performance in current whole-body Ga-PSMA PET-MRI and should be avoided. Increased image noise and imperfections in generated attenuation maps were identified as a paramount cause for image degradation.

Citing Articles

Improved Positron Emission Tomography Quantification: Evaluation of a Maximum-Likelihood Scatter Scaling Algorithm.

Overbeck N, Ahangari S, Conti M, Panin V, Azam A, Kurbegovic S Diagnostics (Basel). 2024; 14(11).

PMID: 38893602 PMC: 11172368. DOI: 10.3390/diagnostics14111075.


Negative predictive value of PSMA PET scan for lymph node staging in patients undergoing robotic radical prostatectomy and pelvic lymph node dissection.

Adiyat K, Pooleri G, Cherian D, Gopalan Nair Santhamma S, Ravichandran K, Sundaram S Int Urol Nephrol. 2023; 55(6):1453-1457.

PMID: 37086333 DOI: 10.1007/s11255-023-03595-0.


Normal Skeletal Standardized Uptake Values Obtained from Quantitative Single-Photon Emission Computed Tomography/Computed Tomography: Time-Dependent Study on Breast Cancer Patients.

Nautiyal A, Jha A, Mithun S, Sawant V, Jadhav R, Khairnar K Indian J Nucl Med. 2022; 36(4):398-411.

PMID: 35125758 PMC: 8771070. DOI: 10.4103/ijnm.ijnm_47_21.


PET/MRI in prostate cancer: a systematic review and meta-analysis.

Evangelista L, Zattoni F, Cassarino G, Artioli P, Cecchin D, Dal Moro F Eur J Nucl Med Mol Imaging. 2020; 48(3):859-873.

PMID: 32901351 PMC: 8036222. DOI: 10.1007/s00259-020-05025-0.


Ga-PSMA-11 PET/MRI: determining ideal acquisition times to reduce noise and increase image quality.

Murthy V, Smith R, Tao D, Lawhn-Heath C, Korenchan D, Larson P EJNMMI Phys. 2020; 7(1):54.

PMID: 32844310 PMC: 7447708. DOI: 10.1186/s40658-020-00322-x.


References
1.
Akamatsu G, Ikari Y, Nishida H, Nishio T, Ohnishi A, Maebatake A . Influence of Statistical Fluctuation on Reproducibility and Accuracy of SUVmax and SUVpeak: A Phantom Study. J Nucl Med Technol. 2015; 43(3):222-6. DOI: 10.2967/jnmt.115.161745. View

2.
Paulus D, Quick H, Geppert C, Fenchel M, Zhan Y, Hermosillo G . Whole-Body PET/MR Imaging: Quantitative Evaluation of a Novel Model-Based MR Attenuation Correction Method Including Bone. J Nucl Med. 2015; 56(7):1061-6. PMC: 4894503. DOI: 10.2967/jnumed.115.156000. View

3.
Partovi S, Kohan A, Gaeta C, Rubbert C, Vercher-Conejero J, Jones R . Image quality assessment of automatic three-segment MR attenuation correction vs. CT attenuation correction. Am J Nucl Med Mol Imaging. 2013; 3(3):291-9. PMC: 3627525. View

4.
Buther F, Vehren T, Schafers K, Schafers M . Impact of Data-driven Respiratory Gating in Clinical PET. Radiology. 2016; 281(1):229-38. DOI: 10.1148/radiol.2016152067. View

5.
Gaertner F, Beer A, Souvatzoglou M, Eiber M, Furst S, Ziegler S . Evaluation of feasibility and image quality of 68Ga-DOTATOC positron emission tomography/magnetic resonance in comparison with positron emission tomography/computed tomography in patients with neuroendocrine tumors. Invest Radiol. 2013; 48(5):263-72. DOI: 10.1097/RLI.0b013e31828234d0. View