Simultaneous Ramp Right Heart Catheterization and Echocardiography in a ReliantHeart Left Ventricular Assist Device

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Many clinicians caring for patients with continuous flow left ventricular assist devices (CF-LVAD) use ramp right heart catheterization (RHC) studies to optimize pump speed and also to troubleshoot CF-LVAD malfunction. An investigational device, the ReliantHeart Heart Assist 5 (Houston, TX), provides the added benefit of an ultrasonic flow probe on the outflow graft that directly measures flow through the CF-LVAD. We performed a simultaneous ramp RHC and echocardiogram on a patient who received the above CF-LVAD to optimize pump parameters and investigate elevated flow through the CF-LVAD as measured by the flow probe. We found that the patient's hemodynamics were optimized at their baseline pump speed, and that the measured cardiac output the Fick principle was lower than that measured by the flow probe. Right heart catheterization may be useful to investigate discrepancies between flow measured by a CF-LVAD and a patient's clinical presentation, particularly in investigational devices where little clinical experience exists. More data is needed to elucidate the correlation between the flow measured by an ultrasonic probe and cardiac output as measured by RHC.

References
1.
Dean D, Jia C, Cabreriza S, DAlessandro D, Dickstein M, Sardo M . Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements. ASAIO J. 1996; 42(5):M671-6. DOI: 10.1097/00002480-199609000-00072. View

2.
Yu L, Kim B, Meng E . Chronically implanted pressure sensors: challenges and state of the field. Sensors (Basel). 2014; 14(11):20620-44. PMC: 4279503. DOI: 10.3390/s141120620. View

3.
Uriel N, Levin A, Sayer G, Mody K, Thomas S, Adatya S . Left Ventricular Decompression During Speed Optimization Ramps in Patients Supported by Continuous-Flow Left Ventricular Assist Devices: Device-Specific Performance Characteristics and Impact on Diagnostic Algorithms. J Card Fail. 2015; 21(10):785-91. DOI: 10.1016/j.cardfail.2015.06.010. View

4.
Uriel N, Sayer G, Addetia K, Fedson S, Kim G, Rodgers D . Hemodynamic Ramp Tests in Patients With Left Ventricular Assist Devices. JACC Heart Fail. 2016; 4(3):208-17. DOI: 10.1016/j.jchf.2015.10.001. View

5.
Salerno C, Sundareswaran K, Schleeter T, Moanie S, Farrar D, Walsh M . Early elevations in pump power with the HeartMate II left ventricular assist device do not predict late adverse events. J Heart Lung Transplant. 2014; 33(8):809-15. DOI: 10.1016/j.healun.2014.02.024. View