» Articles » PMID: 28159069

Longitudinal Assessment of Reprocessing Effectiveness for Colonoscopes and Gastroscopes: Results of Visual Inspections, Biochemical Markers, and Microbial Cultures

Overview
Date 2017 Feb 5
PMID 28159069
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Flexible endoscopes are currently reused following cleaning and high-level disinfection. Contamination has been found on endoscopes, and infections have been linked to gastrointestinal, respiratory, and urologic endoscopes.

Methods: This longitudinal study involved visual inspections with a borescope, microbial cultures, and biochemical tests for protein and adenosine triphosphate to identify endoscopes in need of further cleaning or maintenance. Three assessments were conducted over a 7-month period. Control group endoscopes reprocessed using customary practices were compared with intervention group endoscopes subjected to more rigorous reprocessing.

Results: At final assessment, all endoscopes (N = 20) had visible irregularities. Researchers observed fluid (95%), discoloration, and debris in channels. Of 12 (60%) endoscopes with microbial growth, 4 had no growth until after 48 hours. There were no significant differences in culture results by study group, assessment period, or endoscope type. Similar proportions of control and intervention endoscopes (~20%) exceeded postcleaning biochemical test benchmarks. Adenosine triphosphate levels were higher for gastroscopes than colonoscopes (P = .014). Eighty-five percent of endoscopes required repair due to findings.

Conclusions: More rigorous reprocessing was not consistently effective. Seven-day incubation allowed identification of slow-growing microbes. These findings bolster the need for routine visual inspection and cleaning verification tests recommended in new reprocessing guidelines.

Citing Articles

It's All About the Bubbles: Assessing the Effects of Simethicone on Safety and Efficacy During Colonoscopy.

Perrin M, Nguyen Wenker T, Larson S Dig Dis Sci. 2025; .

PMID: 39994087 DOI: 10.1007/s10620-025-08938-8.


Beyond Endoscopes: Pilot Study of Surgical Instrument Lumen Inspection.

Hopkins K, Adams S, Lamb L, Smart A, Ofstead C Biomed Instrum Technol. 2024; 58(1):25-33.

PMID: 38377297 PMC: 10878677. DOI: 10.2345/0899-8205-58.1.25.


Effect of a disposable endoscope precleaning kit in the cleaning procedure of gastrointestinal endoscope: A multi-center observational study.

Wang Y, Wu Y, Liu X, Li J, Zhan Y, Liu B World J Gastrointest Endosc. 2024; 15(12):705-714.

PMID: 38187912 PMC: 10768042. DOI: 10.4253/wjge.v15.i12.705.


Fluorescent Probe Combined with Photoelectric Analysis Technology for Detection of .

Cui Q, Zhong Y, Shang W, Deng F, Wang B, Wu J Biosensors (Basel). 2023; 13(2).

PMID: 36831916 PMC: 9953772. DOI: 10.3390/bios13020150.


Gastrointestinal endoscope contamination rates - elevators are not only to blame: a systematic review and meta-analysis.

Goyal H, Larsen S, Perisetti A, Larsen N, Ockert L, Adamsen S Endosc Int Open. 2022; 10(6):E840-E853.

PMID: 35692921 PMC: 9187382. DOI: 10.1055/a-1795-8883.