Impact of a Pneumatic Tube System Transport on Hemostasis Parameters Measurement: the Experiment of Cochin Universitary Hospital (AP-HP, Paris, France)

Overview
Specialties Biochemistry
Biology
Date 2017 Jan 31
PMID 28132949
Citations 1
Authors
Affiliations
Soon will be listed here.
Abstract

Samples transported by pneumatic tube system are submitted to forces of acceleration and deceleration which can affect laboratory parameters. At Cochin hospital, majority of samples of hemostasis, except for platelets tests, are transported by pneumatic tube system. The objective of this study was to evaluate the impact of a pneumatic tube system (PTS) transport compared to hand-delivered transport on samples and to qualify Cochin hospital PTS according to requirements of standard ISO 15189. A bibliographical study was made and showed that pneumatic tube system particularly influences platelets tests. Four citrate tubes were collected in 5 healthy volunteers in the maternity: 2 tubes were transported by PTS and 2 others were hand-delivered to the laboratory. Five coagulation tests were analyzed: prothrombine time (PT), activated partial thromboplastin time (aPTT), factor (F) V, FVIII and platelet closure time with PFA-100TM collagen/epinephrine. For each volunteer, the results obtained by PTS and by hand-delivered transport were compared with formula usually used for biological analysis retake: 2.8 x standard deviation of reproductibility variation coefficient (SH GTA 01, COFRAC). This study did not show an impact of PTS on PT, aPTT, FV and FVIII. For PFA-100TM collagen/epinephrine, we noted an impact on 2/5 volunteers. These results, in agreement with the literature, led to the conclusion that Cochin hospital PTS is in compliance to transport samples for usual coagulation tests except platelet tests. This study allowed to issue French recommendations for PTS transport of hemostasis tubes qualification available on "Groupe français d'hémostase et thrombose" Web site.

Citing Articles

Comparison of the speed and quality of innovative and traditional pneumatic tube system transport outside of an emergency laboratory.

Yu J, Zhu G, Cui K, Yu D, Bayartaikishigtai D, Chen Z Heliyon. 2024; 10(10):e31511.

PMID: 38826741 PMC: 11141344. DOI: 10.1016/j.heliyon.2024.e31511.