A New Differential Pressure Flow Meter for Measurement of Human Breath Flow: Simulation and Experimental Investigation

Overview
Journal AIChE J
Date 2016 Nov 8
PMID 27818521
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

The development and performance characterization of a new differential pressure-based flow meter for human breath measurements is presented in this article. The device, called a "Confined Pitot Tube," is comprised of a pipe with an elliptically shaped expansion cavity located in the pipe center, and an elliptical disk inside the expansion cavity. The elliptical disk, named Pitot Tube, is exchangeable, and has different diameters, which are smaller than the diameter of the elliptical cavity. The gap between the disk and the cavity allows the flow of human breath to pass through. The disk causes an obstruction in the flow inside the pipe, but the elliptical cavity provides an expansion for the flow to circulate around the disk, decreasing the overall flow resistance. We characterize the new sensor flow experimentally and theoretically, using Comsol Multiphysics software with laminar and turbulent models. We also validate the sensor, using inhalation and exhalation tests and a reference method.

Citing Articles

Indirect Calorimetry in Spontaneously Breathing, Mechanically Ventilated and Extracorporeally Oxygenated Patients: An Engineering Review.

Priem S, Jonckheer J, De Waele E, Stiens J Sensors (Basel). 2023; 23(8).

PMID: 37112483 PMC: 10144739. DOI: 10.3390/s23084143.


Wearable Piezoelectric Airflow Transducers for Human Respiratory and Metabolic Monitoring.

Jin L, Liu Z, Altintas M, Zheng Y, Liu Z, Yao S ACS Sens. 2022; 7(8):2281-2292.

PMID: 35868024 PMC: 9425556. DOI: 10.1021/acssensors.2c00824.

References
1.
Yoshiya I, Shimada Y, Tanaka K . Evaluation of a hot-wire respiratory flowmeter for clinical applicability. J Appl Physiol Respir Environ Exerc Physiol. 1979; 47(5):1131-5. DOI: 10.1152/jappl.1979.47.5.1131. View

2.
Zhao D, Xian X, Terrera M, Krishnan R, Miller D, Bridgeman D . A pocket-sized metabolic analyzer for assessment of resting energy expenditure. Clin Nutr. 2013; 33(2):341-7. DOI: 10.1016/j.clnu.2013.06.001. View

3.
Ilsley A, Hart J, Withers R, Roberts J . Evaluation of five small turbine-type respirometers used in adult anesthesia. J Clin Monit. 1993; 9(3):196-201. DOI: 10.1007/BF01617028. View

4.
Kreit J, Sciurba F . The accuracy of pneumotachograph measurements during mechanical ventilation. Am J Respir Crit Care Med. 1996; 154(4 Pt 1):913-7. DOI: 10.1164/ajrccm.154.4.8887585. View

5.
Yeh M, Adams T, Gardner R, Yanowitz F . Turbine flowmeter vs. Fleisch pneumotachometer: a comparative study for exercise testing. J Appl Physiol (1985). 1987; 63(3):1289-95. DOI: 10.1152/jappl.1987.63.3.1289. View