Dosimetric Comparison of Flattened and Unflattened Beams for Stereotactic Body Radiation Therapy: Impact of the Size of the PTV on Dynamic Conformal Arc and Volumetric Modulated Arc Therapy

Overview
Journal Phys Med
Date 2016 Oct 21
PMID 27756535
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Flattening filter free (FFF) beams are frequently used for stereotactic body radiation therapy with various treatment modalities: conventional static fields, dynamic conformal arc (DCA) or Rapid Arc (RA). The goal of this study was to obtain some criteria to enable a conscious choice of the employment of FFF beams and of the DCA or RA technique, depending on the PTV size.

Methods And Materials: 24 PTVs from 1.52cm to 445.24cm were studied in various sites: virtual phantom, lung and liver. For each PTV, DCA and RA plans were prepared using two flattened (FF) and two unflattened photon beams. Parameters such as conformity index, gradient index, healthy-tissue and organs at risk mean doses, number of monitor units (MU), beam on time (BOT) were used to quantify obtained dose distributions. Friedman tests and Spearman's rank correlation coefficients were also performed.

Results: No significant differences were found between FF and FFF beams for RA regarding conformity and gradient indices. For DCA, 10FFF is less suitable and forward planning becomes more challenging as PTV volume increases. FFF beams provided a better sparing of healthy-tissues except for 10FFF used with DCA. 6FFF was slightly better than 10FFF in terms of healthy-tissue mean doses. FFF beams generated significantly reduced BOTs and increased MUs. These effects were more pronounced for larger volumes and especially for RA plans.

Conclusions: FFF showed better results than FF beams for the considered plans. 10FFF used with DCA should be used with caution for medium and large volumes.

Citing Articles

Treating early-stage centrally-located non-small cell lung cancer with DCAT-SBRT in centers lacking the VMAT technique: a comprehensive study.

Huang Y, Yang J, Song R, Qin T, Yang M, Liu Y Front Oncol. 2024; 14:1431082.

PMID: 39703844 PMC: 11655335. DOI: 10.3389/fonc.2024.1431082.


Stereotactic body radiotherapy for early-stage lung cancer: a systematic review on the choice of photon energy and linac flattened/unflattened beams.

Gill A, Hirst A, Rowshanfarzad P, Gill S, Bucknell N, Dass J Radiat Oncol. 2024; 19(1):1.

PMID: 38167095 PMC: 10762943. DOI: 10.1186/s13014-023-02392-4.


Stereotactic body radiotherapy with volumetric intensity-modulated arc therapy and flattening filter-free beams: dosimetric considerations.

Brekner M, Imhoff D, Rodel C, Filmann N, Licher J, Ramm U Strahlenther Onkol. 2023; 200(4):346-357.

PMID: 38092967 PMC: 10965745. DOI: 10.1007/s00066-023-02181-8.


A single centre evaluation of the 2019 UK SABR consortium guidelines for primary lung cancer: correlation between Prescription Dose Spillage and inverse Paddick Conformity Index.

Gray S, Kordolaimi S, Norris R, Yiannakis D Rep Pract Oncol Radiother. 2022; 27(2):209-214.

PMID: 36299394 PMC: 9591033. DOI: 10.5603/RPOR.a2021.0135.


Comparison of Moderate Hypofractionated Volumetric-Modulated Arc Therapy Plans With and Without Flattening Filter for Localized Prostate Cancer.

Takakusagi Y, Usui K, Mizoguchi N, Nagatsuka J, Hikage T, Kodama Y Cureus. 2021; 13(9):e18034.

PMID: 34671522 PMC: 8520568. DOI: 10.7759/cureus.18034.