» Articles » PMID: 27707438

Validation of a Wireless, Self-Application, Ambulatory Electroencephalographic Sleep Monitoring Device in Healthy Volunteers

Overview
Specialties Neurology
Psychiatry
Date 2016 Oct 7
PMID 27707438
Citations 36
Authors
Affiliations
Soon will be listed here.
Abstract

Study Objectives: To evaluate the validity of an ambulatory electroencephalographic (EEG) monitor for the estimation of sleep continuity and architecture in healthy adults.

Methods: Healthy, good sleeping participants (n = 14) were fit with both an ambulatory EEG monitor (Sleep Profiler) and a full polysomnography (PSG) montage. EEG recordings were gathered from both devices on the same night, during which sleep was permitted uninterrupted for eight hours. The study was set in an inpatient clinical research suite. PSG and Sleep Profiler records were scored by a neurologist board certified in sleep medicine, blinded to record identification. Agreement between the scored PSG record, the physician-scored Sleep Profiler record, and the Sleep Profiler record scored by an automatic algorithm was evaluated for each sleep stage, with the PSG record serving as the reference.

Results: Results indicated strong percent agreement across stages. Kappa was strongest for Stage N3 and REM. Specificity was high for all stages; sensitivity was low for Wake and Stage N1, and high for Stage N2, Stage N3, and REM. Agreement indices improved for the manually scored Sleep Profiler record relative to the autoscore record.

Conclusions: Overall, the Sleep Profiler yields an EEG record with comparable sleep architecture estimates to PSG. Future studies should evaluate agreement between devices with a clinical sample that has greater periods of wake in order to better understand utility of this device for estimating sleep continuity indices, such as sleep onset latency and wake after sleep onset.

Citing Articles

Use of a wearable device to improve sleep quality.

Moore S, Carey E, Finikiotis K, Ford K, Zane R, Green K Front Digit Health. 2025; 6:1384173.

PMID: 39968357 PMC: 11834000. DOI: 10.3389/fdgth.2024.1384173.


Mobile Sleep Lab: Comparison of polysomnographic parameters with a conventional sleep laboratory.

Suzuki C, Suzuki Y, Abe T, Kanbayashi T, Fukusumi S, Kokubo T PLoS One. 2025; 20(1):e0316579.

PMID: 39775303 PMC: 11706495. DOI: 10.1371/journal.pone.0316579.


Optimizing wearable single-channel electroencephalography sleep staging in a heterogeneous sleep-disordered population using transfer learning.

van der Aar J, van Gilst M, van den Ende D, Fonseca P, van Wetten B, Janssen H J Clin Sleep Med. 2024; 21(2):315-323.

PMID: 39347545 PMC: 11789263. DOI: 10.5664/jcsm.11380.


Performance Investigation of Somfit Sleep Staging Algorithm.

McMahon M, Goldin J, Kealy E, Wicks D, Zilberg E, Freeman W Nat Sci Sleep. 2024; 16:1027-1043.

PMID: 39071546 PMC: 11277903. DOI: 10.2147/NSS.S463026.


Influence of Obstructive Apnea Index on Persistent Left Ventricular Dysfunction in Patients with ST-Segment Elevation Myocardial Infarction.

Kirigaya J, Iwahashi N, Ishigami T, Abe T, Gohbara M, Hanajima Y J Clin Med. 2024; 13(4).

PMID: 38398299 PMC: 10888575. DOI: 10.3390/jcm13040986.


References
1.
Collop N, Anderson W, Boehlecke B, Claman D, Goldberg R, Gottlieb D . Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2008; 3(7):737-47. PMC: 2556918. View

2.
Gujar N, Yoo S, Hu P, Walker M . Sleep deprivation amplifies reactivity of brain reward networks, biasing the appraisal of positive emotional experiences. J Neurosci. 2011; 31(12):4466-74. PMC: 3086142. DOI: 10.1523/JNEUROSCI.3220-10.2011. View

3.
Armitage R, Hoffmann R, Trivedi M, Rush A . Slow-wave activity in NREM sleep: sex and age effects in depressed outpatients and healthy controls. Psychiatry Res. 2000; 95(3):201-13. DOI: 10.1016/s0165-1781(00)00178-5. View

4.
AGNEW Jr H, WEBB W, Williams R . The first night effect: an EEG study of sleep. Psychophysiology. 1966; 2(3):263-6. DOI: 10.1111/j.1469-8986.1966.tb02650.x. View

5.
Stepnowsky C, Levendowski D, Popovic D, Ayappa I, Rapoport D . Scoring accuracy of automated sleep staging from a bipolar electroocular recording compared to manual scoring by multiple raters. Sleep Med. 2013; 14(11):1199-207. DOI: 10.1016/j.sleep.2013.04.022. View