Film Based Verification of Calculation Algorithms Used for Brachytherapy Planning-getting Ready for Upcoming Challenges of MBDCA

Overview
Date 2016 Sep 21
PMID 27648087
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Well-known defect of TG-43 based algorithms used in brachytherapy is a lack of information about interaction cross-sections, which are determined not only by electron density but also by atomic number. TG-186 recommendations with using of MBDCA (model-based dose calculation algorithm), accurate tissues segmentation, and the structure's elemental composition continue to create difficulties in brachytherapy dosimetry. For the clinical use of new algorithms, it is necessary to introduce reliable and repeatable methods of treatment planning systems (TPS) verification. The aim of this study is the verification of calculation algorithm used in TPS for shielded vaginal applicators as well as developing verification procedures for current and further use, based on the film dosimetry method.

Material And Methods: Calibration data was collected by separately irradiating 14 sheets of Gafchromic(®) EBT films with the doses from 0.25 Gy to 8.0 Gy using HDR (192)Ir source. Standard vaginal cylinders of three diameters were used in the water phantom. Measurements were performed without any shields and with three shields combination. Gamma analyses were performed using the VeriSoft(®) package.

Results: Calibration curve was determined as third-degree polynomial type. For all used diameters of unshielded cylinder and for all shields combinations, Gamma analysis were performed and showed that over 90% of analyzed points meets Gamma criteria (3%, 3 mm).

Conclusions: Gamma analysis showed good agreement between dose distributions calculated using TPS and measured by Gafchromic films, thus showing the viability of using film dosimetry in brachytherapy.

Citing Articles

Quantitative study on dose distribution of Freiburg flap for keloid high-dose-rate brachytherapy based on MatriXX.

Ni J, Gan G, Xu X J Appl Clin Med Phys. 2023; 24(9):e14118.

PMID: 37593834 PMC: 10476986. DOI: 10.1002/acm2.14118.


Advanced dose calculation algorithm in superficial brachytherapy - the impact of tissue inhomogeneity on treatment plan dosimetry.

Szlag M, Kellas-Sleczka S, Wojcieszek P, Stankiewicz M, Cholewka A, Pruefer A J Contemp Brachytherapy. 2021; 13(4):441-446.

PMID: 34484359 PMC: 8407266. DOI: 10.5114/jcb.2021.106541.


Quality Assurance Procedures based on Dosimetric, Gamma Analysis as a Fast Reliable Tool for Commissioning Brachytherapy Treatment Planning Systems.

Zwierzchowski G, Bieleda G, Skowronek J Radiol Oncol. 2018; 51(4):469-474.

PMID: 29333127 PMC: 5765325. DOI: 10.1515/raon-2017-0050.

References
1.
Chiu-Tsao S, Ho Y, Shankar R, Wang L, Harrison L . Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies. Med Phys. 2005; 32(11):3350-4. DOI: 10.1118/1.2065467. View

2.
Zourari K, Pantelis E, Moutsatsos A, Sakelliou L, Georgiou E, Karaiskos P . Dosimetric accuracy of a deterministic radiation transport based (192)Ir brachytherapy treatment planning system. Part III. Comparison to Monte Carlo simulation in voxelized anatomical computational models. Med Phys. 2013; 40(1):011712. DOI: 10.1118/1.4770275. View

3.
Granero D, Vijande J, Ballester F, Rivard M . Dosimetry revisited for the HDR 192Ir brachytherapy source model mHDR-v2. Med Phys. 2011; 38(1):487-94. DOI: 10.1118/1.3531973. View

4.
Palmer A, Bradley D, Nisbet A . Physics-aspects of dose accuracy in high dose rate (HDR) brachytherapy: source dosimetry, treatment planning, equipment performance and in vivo verification techniques. J Contemp Brachytherapy. 2013; 4(2):81-91. PMC: 3552629. DOI: 10.5114/jcb.2012.29364. View

5.
Aldelaijan S, Mohammed H, Tomic N, Liang L, DeBlois F, Sarfehnia A . Radiochromic film dosimetry of HDR (192)Ir source radiation fields. Med Phys. 2011; 38(11):6074-83. DOI: 10.1118/1.3651482. View