» Articles » PMID: 27606263

Phantom Study of a New Laser-Etched Needle for Improving Visibility During Ultrasonography-Guided Lumbar Medial Branch Access With Novices

Overview
Journal Ann Rehabil Med
Date 2016 Sep 9
PMID 27606263
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare the visibility and procedural parameters between a standard spinal needle and a new laser-etched needle (LEN) in real-time ultrasonography guided lumbar medial branch access in a phantom of the lumbosacral spine.

Methods: We conducted a prospective single-blinded observational study at a rehabilitation medicine center. A new model of LEN was manufactured with a standard 22-gauge spinal needle and a laser etching machine. Thirty-two inexperienced polyclinic medical students performed ultrasonography-guided lumbar medial branch access using both a standard spinal needle and a LEN with scanning protocol. The outcomes included needle visibility score, needle elapsed time, first-pass success rate, and number of needle sticks.

Results: The LEN received significantly better visibility scores and shorter needle elapsed time compared to the standard spinal needle. First-pass success rate and the number of needle sticks were not significantly different between needles.

Conclusion: A new LEN is expected to offer better visibility and enable inexperienced users to perform an ultrasonography-guided lumbar medial branch block more quickly. However, further study of variables may be necessary for clinical application.

Citing Articles

Ultrasound-guided needle tracking with deep learning: A novel approach with photoacoustic ground truth.

Hui X, Rajendran P, Ling T, Dai X, Xing L, Pramanik M Photoacoustics. 2024; 34:100575.

PMID: 38174105 PMC: 10761306. DOI: 10.1016/j.pacs.2023.100575.


Autonomous Spinal Robotic System for Transforaminal Lumbar Epidural Injections: A Proof of Concept of Study.

Margalit A, Phalen H, Gao C, Ma J, Suresh K, Jain P Global Spine J. 2022; 14(1):138-145.

PMID: 35467447 PMC: 10676186. DOI: 10.1177/21925682221096625.


Real-time ultrasound-computed tomography image fusion for transforaminal lumbar approach: a lumbosacral spine phantoms study.

Emmanuel G, Arnaud P, Jonathan G, Marie R, Iulia M, Sanjiva P Eur Spine J. 2021; 30(5):1270-1276.

PMID: 33713176 DOI: 10.1007/s00586-021-06794-4.


Efficacy of using a 3D printed lumbosacral spine phantom in improving trainee proficiency and confidence in CT-guided spine procedures.

Li Y, Li Z, Ammanuel S, Gillan D, Shah V 3D Print Med. 2019; 4(1):7.

PMID: 30649653 PMC: 6179970. DOI: 10.1186/s41205-018-0031-x.

References
1.
Culp W, McCowan T, Goertzen T, Habbe T, Hummel M, Leveen R . Relative ultrasonographic echogenicity of standard, dimpled, and polymeric-coated needles. J Vasc Interv Radiol. 2000; 11(3):351-8. DOI: 10.1016/s1051-0443(07)61429-8. View

2.
Gottlieb R, Robinette W, Rubens D, Hartley D, Fultz P, Violante M . Coating agent permits improved visualization of biopsy needles during sonography. AJR Am J Roentgenol. 1998; 171(5):1301-2. DOI: 10.2214/ajr.171.5.9798867. View

3.
Chin K, Perlas A, Chan V, Brull R . Needle visualization in ultrasound-guided regional anesthesia: challenges and solutions. Reg Anesth Pain Med. 2009; 33(6):532-44. DOI: 10.1016/j.rapm.2008.06.002. View

4.
Kaplan M, DREYFUSS P, Halbrook B, Bogduk N . The ability of lumbar medial branch blocks to anesthetize the zygapophysial joint. A physiologic challenge. Spine (Phila Pa 1976). 1998; 23(17):1847-52. DOI: 10.1097/00007632-199809010-00008. View

5.
Marks R, Semple A . Spinal anaesthesia after facet joint injection. Anaesthesia. 1988; 43(1):65-6. DOI: 10.1111/j.1365-2044.1988.tb05442.x. View