Abstract

Purpose: Percutaneous sacroiliac (SI) fixation of unstable posterior pelvic ring injuries is a widely accepted procedure. The complex sacral anatomy with narrow osseous corridors for SI screw placement makes this procedure technically challenging. Techniques are constantly evolving as a result of better understanding of the posterior pelvic anatomy. Recently developed tools include fluoroscopy-based computer-assisted navigation, which can be two-dimensional (2D) or three-dimensional (3D). Our goal is to determine the relevant technical considerations and clinical outcomes associated with these modalities by reviewing the published research. We hypothesize that 3D fluoroscopy-based navigation is safer and superior to its 2D predecessor with respect to lower radiation dose and more accurate SI screw placement.

Methods: We searched four medical databases to identify English-language studies of 2D and 3D fluoroscopy-based navigation from January 1990 through August 2015. We included articles reporting imaging techniques and outcomes of closed posterior pelvic ring fixation with percutaneous SI screw fixation. Injuries included in the study were sacral fractures (52 patients), sacroiliac fractures (88 patients), lateral compression fractures (20 patients), and anteroposterior compression type pelvic fractures (8 patients). We excluded articles on open reduction of posterior pelvic ring injuries and solely anatomic studies. We then reviewed these studies for technical considerations and outcomes associated with these technologies.

Results: Six studies were included in our analysis. Results of these studies indicate that 3D fluoroscopy-based navigation is associated with a lower radiation dose and lower rate of screw malpositioning compared with 2D fluoroscopy-based systems.

Conclusions: It may be advantageous to combine modern imaging modalities such as 3D fluoroscopy with computer-assisted navigation for percutaneous screw fixation in the posterior pelvis.

Citing Articles

Accuracy in navigated percutaneous sacroiliac screw fixation: a systematic review and meta-analysis.

Haveman R, Buchmann L, Haefeli P, Beeres F, Babst R, Link B BMC Surg. 2025; 25(1):89.

PMID: 40045283 PMC: 11881291. DOI: 10.1186/s12893-025-02813-z.


A Projective-Geometry-Aware Network for 3D Vertebra Localization in Calibrated Biplanar X-Ray Images.

Ye K, Sun W, Tao R, Zheng G Sensors (Basel). 2025; 25(4).

PMID: 40006352 PMC: 11858964. DOI: 10.3390/s25041123.


Calculation of CT ideal screw path and safety angle before percutaneous sacroiliac screw placement.

Zhang Z, Chen J, Chen X, Sun R Arch Orthop Trauma Surg. 2025; 145(1):153.

PMID: 39891735 PMC: 11787272. DOI: 10.1007/s00402-025-05774-3.


Preoperative 3D printing planning technology combined with orthopedic surgical robot-assisted minimally invasive screw fixation for the treatment of pelvic fractures: a retrospective study.

Jing Y, Chang L, Cong B, Wang J, Chen M, Tang Z PeerJ. 2024; 12:e18632.

PMID: 39677955 PMC: 11646416. DOI: 10.7717/peerj.18632.


Percutaneous sacroiliac screw fixation with a 3D robot-assisted image-guided navigation system : Technical solutions.

Link B, Haveman R, van de Wall B, Baumgartner R, Babst R, Beeres F Oper Orthop Traumatol. 2024; 37(1):3-13.

PMID: 39556213 PMC: 11790701. DOI: 10.1007/s00064-024-00871-9.


References
1.
Papanastassiou I, Setzer M, Eleraky M, Baaj A, Nam T, Binitie O . Minimally invasive sacroiliac fixation in oncologic patients with sacral insufficiency fractures using a fluoroscopy-based navigation system. J Spinal Disord Tech. 2010; 24(2):76-82. DOI: 10.1097/BSD.0b013e3181df8e6b. View

2.
Kim J, Oh C, Oh J, Lee H, Min W, Kyung H . Percutaneous iliosacral screwing in pelvic ring injury using three-dimensional fluoroscopy. J Orthop Sci. 2012; 18(1):87-92. DOI: 10.1007/s00776-012-0320-y. View

3.
Peng K, Li Y, Hsu W, Wu M, Yang J, Hsu C . Intraoperative computed tomography with integrated navigation in percutaneous iliosacral screwing. Injury. 2012; 44(2):203-8. DOI: 10.1016/j.injury.2012.09.017. View

4.
Schep N, Haverlag R, Van Vugt A . Computer-assisted versus conventional surgery for insertion of 96 cannulated iliosacral screws in patients with postpartum pelvic pain. J Trauma. 2004; 57(6):1299-302. DOI: 10.1097/01.ta.0000133573.53587.2e. View

5.
Stockle U, Konig B, Schaser K, Melcher I, Haas N . [CT and fluoroscopy based navigation in pelvic surgery]. Unfallchirurg. 2003; 106(11):914-20. DOI: 10.1007/s00113-003-0677-7. View