» Articles » PMID: 27497923

Small Field Correction Factors for the IBA Razor

Overview
Journal Phys Med
Date 2016 Aug 8
PMID 27497923
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

The IBA Razor diode supersedes the IBA SFD and is intended for use in small fields. However, its behaviour in small fields has not yet been quantified. In this work, we examine the response of the Razor diode against the air core scintillation dosimeter (FOD) and Gafchromic film in photon beams from three Varian linac beams. Fields between 4mm and 30mm in width were measured, both with and without a flattening filter and at two energies. The Razor exhibited an over-response of up to 4.5% for MLC collimated fields and 7.1% for stereotactic cone collimated fields. The presence of the flattening filter altered the over-response by up to 1.5%. The small field correction factors are tabulated and agree with the mathematical relation of Liu et al. (2014). Four samples of the Razor were used, two having received a significant prior dose. The correction factors for the four samples differed and may depend on their dose history.

Citing Articles

Determination of correction factors in small MLC-defined fields for the Razor and microSilicon diode detectors and evaluation of the suitability of the IAEA TRS-483 protocol for multiple detectors.

McGrath A, Golmakani S, Williams T J Appl Clin Med Phys. 2022; 23(7):e13657.

PMID: 35652320 PMC: 9278669. DOI: 10.1002/acm2.13657.


Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams.

Gul A, Fukuda S, Mizuno H, Taku N, Kakakhel M, Mirza S J Appl Clin Med Phys. 2020; 21(11):23-36.

PMID: 33078544 PMC: 7700919. DOI: 10.1002/acm2.13007.


Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel.

Casar B, Gershkevitsh E, Mendez I, Jurkovic S, Huq M Med Phys. 2019; 47(1):242-259.

PMID: 31677278 PMC: 7003763. DOI: 10.1002/mp.13894.


"Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Shukaili K, Corde S, Petasecca M, Pereveratylo V, Lerch M, Jackson M J Appl Clin Med Phys. 2018; 19(4):114-124.

PMID: 29790261 PMC: 6036391. DOI: 10.1002/acm2.12345.