» Articles » PMID: 27476640

Suture Anchor Fixation in Osteoporotic Bone: A Biomechanical Study in an Ovine Model

Overview
Journal Arthroscopy
Specialty Orthopedics
Date 2016 Aug 2
PMID 27476640
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the load to failure strength of anchor techniques suitable for osteoporotic bone.

Methods: Using an osteoporotic ovine model in 72 humeri, 6 fixation techniques were tested. Group 1: two interlocking 5-mm anchors with fewer, wider threads; group 2: one 5-mm anchor; group 3: one 5-mm anchor reinforced laterally by a 6.5-mm cancellous screw; group 4: one 5-mm anchor in an area reinforced with a cancellous plug; group 5: two interlocking 5-mm anchors with smaller threads; group 6: one 5-mm smaller threaded anchor. After a 10-N preload, the specimens were cyclically loaded between 10 N and 30 N for 50 cycles and then destructively tested. Peak-to-peak displacement, cyclic elongation, ultimate load, stiffness, and failure mode were recorded.

Results: Group 1 had lower peak-to-peak displacement than group 3 (P = .001), group 5 (P = .001), and group 6 (P = .033). In addition, group 1 showed lower cyclic elongation than group 3 (P = .001), group 5 (P = .035), and group 6 (P = .001). Group 1 had a higher ultimate load than group 2 (P = .002), group 3 (P = .019), and group 6 (P = .006). Group 1 also showed higher stiffness than group 2 (P = .007) and group 3 (P = .022). Mode of failure was predominantly caused by anchor pullout for all of the groups except group 3, which mainly failed by suture rupture.

Conclusions: Two interlocking suture anchors are stronger than a single anchor in osteoporotic bone. The anchor with fewer, wider threads and a smaller core diameter showed greater strength and less elongation than the other constructs. Reinforcement by cancellous autografting increased suture anchor strength.

Clinical Relevance: Considering concerns about suture anchor pullout from osteoporotic bone, interlocking a second suture anchor with the first increases load to failure resistance. Anchors with small core diameters and fewer but wider threads are more efficient in osteoporotic bone.

Citing Articles

Arthroscopic transosseous anchorless rotator cuff repair reduces bone defects related to peri-implant cyst formation: a comparison with conventional suture anchors using propensity score matching.

Jeong H, Lee J, Kim Y, Rhee S, Oh J Clin Shoulder Elb. 2023; 26(3):276-286.

PMID: 37559521 PMC: 10497926. DOI: 10.5397/cise.2023.00052.


An Injectable Calcium Phosphate Bone Graft Substitute Improves the Pullout Strength of Various Suture Anchor Designs in an Osteoporotic Bone Model.

Diaz M, Munassi S, Teytelbaum D, Pipitone A, Baker C Arthrosc Sports Med Rehabil. 2023; 5(2):e423-e433.

PMID: 37101877 PMC: 10123442. DOI: 10.1016/j.asmr.2023.01.010.


Risk Factors and Corresponding Management for Suture Anchor Pullout during Arthroscopic Rotator Cuff Repair.

Li X, Xiao Y, Shu H, Sun X, Nie M J Clin Med. 2022; 11(22).

PMID: 36431347 PMC: 9699024. DOI: 10.3390/jcm11226870.


Development of a Novel Hybrid Suture Anchor for Osteoporosis by Integrating Titanium 3D Printing and Traditional Machining.

Chen C, Chang W, Chen Y, Chen K, Huang S, Hsueh H Int J Bioprint. 2022; 8(4):608.

PMID: 36404776 PMC: 9668590. DOI: 10.18063/ijb.v8i4.608.


Biomechanical properties of a suture anchor system from human allogenic mineralized cortical bone matrix for rotator cuff repair.

Schanda J, Obermayer-Pietsch B, Sommer G, Heuberer P, Laky B, Muschitz C BMC Musculoskelet Disord. 2022; 23(1):422.

PMID: 35513813 PMC: 9069722. DOI: 10.1186/s12891-022-05371-0.