Comparison Between Noninvasive Measurement of Central Venous Pressure Using Near Infrared Spectroscopy with an Invasive Central Venous Pressure Monitoring in Cardiac Surgical Intensive Care Unit

Overview
Specialty Anesthesiology
Date 2016 Jul 12
PMID 27397443
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Central venous pressure (CVP) measurement is essential in the management of certain clinical situations, including cardiac failure, volume overload and sepsis. CVP measurement requires catheterization of the central vein which is invasive and may lead to complications. The aim of this study was to evaluate the accuracy of measurement of CVP using a new noninvasive method based on near infrared spectroscopy (NIRS) in a group of cardiac surgical Intensive Care Unit (ICU) patients.

Methodology: Thirty patients in cardiac surgical ICU were enrolled in the study who had an in situ central venous catheter (CVC). Sixty measurements were recorded in 1 h for each patient. A total of 1800 values were compared between noninvasive CVP (CVPn) obtained from Mespere VENUS 2000 CVP system and invasive CVP (CVPi) obtained from CVC.

Results: Strong positive correlation was found between CVPi and CVPn (R = 0.9272, P < 0.0001). Linear regression equation - CVPi = 0.5404 + 0.8875 × CVPn (r2 = 0.86, P < 0.001), Bland-Altman bias plots showed mean difference ± standard deviation and limits of agreement: -0.31 ± 1.36 and - 2.99 to + 2.37 (CVPi-CVPn).

Conclusion: Noninvasive assessment of the CVP based on NIRS yields readings consistently close to those measured invasively. CVPn may be a clinically useful substitute for CVPi measurements with an advantage of being simple and continuous. It is a promising tool for early management of acute state wherein knowledge of CVP is helpful.

Citing Articles

Haemodynamic monitoring during noncardiac surgery: past, present, and future.

Kouz K, Thiele R, Michard F, Saugel B J Clin Monit Comput. 2024; 38(3):565-580.

PMID: 38687416 PMC: 11164815. DOI: 10.1007/s10877-024-01161-2.


Design and Haemodynamic Analysis of a Novel Anchoring System for Central Venous Pressure Measurement.

Manavi T, Ijaz M, OGrady H, Nagy M, Martina J, Finucane C Sensors (Basel). 2022; 22(21).

PMID: 36366251 PMC: 9659073. DOI: 10.3390/s22218552.


Changing trend of abdominal regional oxygen saturation in very/extremely low birth weight infants in the early postnatal stage: a prospective study.

Zhang J, Guan R, Pan P, Lu W, Zhang H Zhongguo Dang Dai Er Ke Za Zhi. 2021; 23(10):1015-1020.

PMID: 34719416 PMC: 8549648. DOI: 10.7499/j.issn.1008-8830.2105156.


A Novel Non-Invasive Device for the Assessment of Central Venous Pressure in Hospital, Office and Home.

Marcelli E, Cercenelli L, Bortolani B, Marini S, Arfilli L, Capucci A Med Devices (Auckl). 2021; 14:141-154.

PMID: 34012302 PMC: 8128499. DOI: 10.2147/MDER.S307775.


Noninvasive Technique for Measuring Central Venous and Arterial Pressure Using Controlled Compression Sonography.

Tomoeda H, Sawada K, Chihara S Ann Vasc Dis. 2021; 13(4):397-403.

PMID: 33391557 PMC: 7758583. DOI: 10.3400/avd.oa.20-00058.


References
1.
Ward K, Tiba M, Draucker G, Proffitt E, Barbee R, Gunnerson K . A novel noninvasive impedance-based technique for central venous pressure measurement. Shock. 2009; 33(3):269-73. DOI: 10.1097/SHK.0b013e3181ab9b9b. View

2.
Dellinger R, Levy M, Carlet J, Bion J, Parker M, Jaeschke R . Surviving Sepsis Campaign: international guidelines for management of severe sepsis and septic shock: 2008. Crit Care Med. 2007; 36(1):296-327. DOI: 10.1097/01.CCM.0000298158.12101.41. View

3.
Briscoe C . A comparison of jugular and central venous pressure measurements during anaesthesia. Br J Anaesth. 1973; 45(2):173-8. DOI: 10.1093/bja/45.2.173. View

4.
Charalambous C, Barker T, Zipitis C, Siddique I, Swindell R, Jackson R . Comparison of peripheral and central venous pressures in critically Ill patients. Anaesth Intensive Care. 2003; 31(1):34-9. DOI: 10.1177/0310057X0303100106. View

5.
Marcelino P, Fernandes A, Marum S, Ribeiro J . Non-invasive evaluation of central venous pressure by echocardiography. Rev Port Cardiol. 2002; 21(2):125-33. View