Background: Balancing is an important part of a total knee procedure, and in recent years, more emphasis has been given to quantifying the process.

Methods: During 101 total knee surgeries, initial bone cuts were made using navigation. Lateral and medial contact forces were determined throughout flexion using an instrumented tibial trial. Balancing was defined as a ratio of the medial and total force, the target being 0.5 (equal lateral and medial forces). Based on the initial values, surgical corrections were selected to achieve balancing. The most common corrections were soft tissue releases (63 incidences), including MCL, posterolateral corner, posteromedial corner, and changing tibial insert thicknesses (34 incidences).

Results: After final balancing, the mean ratio was 0.52 ± 0.14, between 0.35 and 0.65 being achieved in 80% of cases. In 84% of cases, only 0-2 corrections were required. The average total force on the condyles was 215 ± 86 N.

Conclusion: Our study provides data to surgeons on the results to expect when balancing a knee, which can enhance both accuracy and consistency of the procedure.

Citing Articles

Predictive Gap-balancing Reduces the Extent of Soft-tissue Adjustment Required After Bony Resection in Robot-assisted Total Knee Arthroplasty-A Comparison With Simulated Measured Resection.

Orsi A, Wakelin E, Plaskos C, Gupta S, Sullivan J Arthroplast Today. 2022; 16:1-8.

PMID: 35620585 PMC: 9126743. DOI: 10.1016/j.artd.2022.03.025.


Total Knee Arthroplasty and Intra-Articular Pressure Sensors: Can They Assist Surgeons with Intra-Operative Decisions?.

Yapp L, Robinson P, Clement N, Scott C Curr Rev Musculoskelet Med. 2021; 14(6):361-368.

PMID: 34962638 PMC: 8733123. DOI: 10.1007/s12178-021-09724-5.


Current role of intraoperative sensing technology in total knee arthroplasty.

Batailler C, Swan J, Marinier E, Servien E, Lustig S Arch Orthop Trauma Surg. 2021; 141(12):2255-2265.

PMID: 34427757 DOI: 10.1007/s00402-021-04130-5.


Can Intraoperative Intra-Articular Loads Predict Postoperative Knee Joint Laxity Following Total Knee Arthroplasty? A Cadaver Study with Smart Tibial Trays.

Shah D, Taylan O, Verstraete M, Berger P, Vandenneucker H, Scheys L Sensors (Basel). 2021; 21(15).

PMID: 34372312 PMC: 8347859. DOI: 10.3390/s21155078.


How to Quantitatively Balance a Total Knee? A Surgical Algorithm to Assure Balance and Control Alignment.

Moore R, Conditt M, Roche M, Verstraete M Sensors (Basel). 2021; 21(3).

PMID: 33498576 PMC: 7864328. DOI: 10.3390/s21030700.