Laboratory Assessment of Thermal Characteristics of Three Phacoemulsification Tip Designs Operated Using Torsional Ultrasound

Overview
Journal Clin Ophthalmol
Publisher Dove Medical Press
Specialty Ophthalmology
Date 2016 Jul 1
PMID 27358554
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Ultrasound activation of phacoemulsification (phaco) tips can create considerable thermal energy that may increase the risk of tissue damage during cataract surgery. The purpose of this study was to define the thermal profiles of three phaco tip designs in simulated surgical conditions.

Methods: In this laboratory investigation, sleeved phaco tips (mini-flared Kelman(®) tip with aspiration bypass port and Intrepid(®) Balanced Tip with aspiration bypass port, and MST A1 bent-mini phaco tip (without aspiration bypass) were tested using an ultrasonic phaco device operated in torsional mode at power levels of 50%, 75%, and 100% amplitude. An automated fixture applied a 30 g load to simulate compression against the incision site, leading to friction between the silicone sleeve and the titanium tip. Temperature was recorded by high rate infrared imaging under conditions of free flow and occlusion, which was simulated by clamping the aspiration line. Data were summarized using descriptive statistics.

Results: Baseline temperatures of ~26°C were observed for all tips. During ultrasonic operation at 50%, 75%, and 100% amplitude, temperatures were lower for the mini-flared and balanced tips versus the bent-mini tip, both when load was applied and during occlusion. The bent-mini tip reached temperatures as high as 70°C during occlusion with load when operated at 100% amplitude, whereas the mini-flared tip remained <50°C, and the balanced tip remained <36°C in all test conditions. For the mini-flared and balanced tips, temperature increases during operation were not markedly different from free flow and no-load conditions when occlusion or frictional events were simulated.

Conclusion: In all experiments for each tip design, increasing ultrasound power was associated with greater increases in tip temperature. Tip temperatures increased with applied load, but marked temperature increases during occlusion were observed only with the bent-mini tip. The balanced tip produced minimal thermal peaks in all tests.

Citing Articles

In vitro analysis of clinically relevant aspects of a polymer-coated phacoemulsification tip vs a traditional tip.

Zacharias J J Cataract Refract Surg. 2023; 49(12):1264-1269.

PMID: 37464557 PMC: 10664810. DOI: 10.1097/j.jcrs.0000000000001267.


Temperature Change of Ophthalmic Viscosurgical Devices in a Bi-Chamber Set-Up at a Flow of 0 and 20mL/min.

Jensen N, Ungricht E, Harris J, Zaugg B, Barlow W, Murri M Clin Ophthalmol. 2023; 17:555-560.

PMID: 36798796 PMC: 9926926. DOI: 10.2147/OPTH.S389136.


Eyelid and scleral thermal injury following phacoemulsification in silicone oil: a case report.

Lee Y, Chen S, Hung J BMC Ophthalmol. 2022; 22(1):420.

PMID: 36333675 PMC: 9636749. DOI: 10.1186/s12886-022-02646-x.


Comparisons of the energy efficiency and intraocular safety of two torsional phacoemulsification tips.

Sun Y, Cao R, Liu Z, Xia H, Cen Y, Gao L BMC Ophthalmol. 2022; 22(1):392.

PMID: 36192712 PMC: 9528131. DOI: 10.1186/s12886-022-02619-0.

References
1.
Khodabakhsh A, Zaidman G, Tabin G . Corneal surgery for severe phacoemulsification burns. Ophthalmology. 2004; 111(2):332-4. DOI: 10.1016/j.ophtha.2003.06.004. View

2.
Sugar A, Schertzer R . Clinical course of phacoemulsification wound burns. J Cataract Refract Surg. 1999; 25(5):688-92. DOI: 10.1016/s0886-3350(99)00021-8. View

3.
Tsinopoulos I, Lamprogiannis L, Tsaousis K, Mataftsi A, Symeonidis C, Chalvatzis N . Surgical outcomes in phacoemulsification after application of a risk stratification system. Clin Ophthalmol. 2013; 7:895-9. PMC: 3663436. DOI: 10.2147/OPTH.S42726. View

4.
Osher R, Injev V . Thermal study of bare tips with various system parameters and incision sizes. J Cataract Refract Surg. 2006; 32(5):867-72. DOI: 10.1016/j.jcrs.2005.06.054. View

5.
Olson M, Miller K . In-air thermal imaging comparison of Legacy AdvanTec, Millennium, and Sovereign WhiteStar phacoemulsification systems. J Cataract Refract Surg. 2005; 31(8):1640-7. DOI: 10.1016/j.jcrs.2005.01.016. View