Background:
Coronary CT angiography (CTA) is gaining widespread acceptance for the non-invasive evaluation of coronary arteries. However, radiation exposure and administration of iodinated contrast agents are still reasons of some concern. The 80 kV tube voltage increases the attenuation of iodine, allowing to use lower iodine concentration contrast agents for coronary CTA.
Objective:
We evaluated the diagnostic accuracy of coronary CTA performed with 64-slice scanner, 80 kV tube voltage, iterative reconstruction algorithm and ultra-low concentration contrast medium to reduce iodine load and radiation dose.
Methods:
We enrolled 45 patients with low body weight and indication for elective invasive coronary angiography (ICA). All patients received an 80 ml bolus of Iodixanol-270 at an infusion rate of 5 mL/s and underwent coronary CTA (80 kV and 500-550 mA) with prospective ECG-triggering. Image quality score, type of artifacts, coronary CTA evaluability, diagnostic accuracy and radiation exposure were assessed.
Results:
Pre-test probability of CAD was low-to-intermediate (48%). Accordingly, the prevalence of obstructive CAD was 47% (21 out of 45 patients). Most (93%) of the patients were pre-treated with intravenous metoprolol before scanning and achieved a heart rate suitable for prospective ECG-triggering coronary CTA (53 ± 3 bpm). The mean effective dose and iodine load were 1.1 ± 0.4 mSv and 21.6 gI, respectively. We rated 443 out of 720 coronary segments as being of excellent image quality. In a segment-based model, coronary evaluability (number of coronary segments evaluable/total number of coronary segments), was 97% (699/720 segments). In a segment-based analysis, sensitivity, specificity, positive predictive value, negative predictive value and accuracy for >50% coronary stenosis identification vs. ICA were 89%, 99%, 89%, 99% and 99%, respectively. In a patient-based analysis, sensitivity, specificity, positive predictive value, negative predictive value and accuracy were 94%, 89%, 83%, 96% and 91%, respectively.
Conclusions:
In patients with low body weight, image quality and diagnostic accuracy of ultra-low radiation dose and low-iodine load coronary CTA are good and similar to values reported in the literature for standard tube voltage and iodine load protocols.
Citing Articles
Efficacy and safety of 128 slice CT low-dose scanning technique combined with low-density contrast agent in computed tomography angiography of diabetic foot and lower limb.
Gao Y, Ma J, Lu T, Qi J
Am J Transl Res. 2024; 16(11):7136-7144.
PMID: 39678539
PMC: 11645635.
DOI: 10.62347/VAFX9854.
Refining the radiation and contrast medium dose in weight-grouped scanning protocols for coronary CT angiography.
Zhang Y, Wang Y, Li J, Zhang G, Di A, Yuan H
J Appl Clin Med Phys. 2023; 24(7):e14041.
PMID: 37211752
PMC: 10338851.
DOI: 10.1002/acm2.14041.
Low dose contrast media in step-and-shoot coronary angiography with third-generation dual-source computed tomography: feasibility of using 30 mL of contrast media in patients with body surface area <1.7 m.
Jin L, Jie B, Gao Y, Jiang A, Weng T, Li M
Quant Imaging Med Surg. 2021; 11(6):2598-2609.
PMID: 34079726
PMC: 8107323.
DOI: 10.21037/qims-20-500.
Right Ventricular Mass Quantification Using Cardiac CT and a Semiautomatic Three-Dimensional Hybrid Segmentation Approach: A Pilot Study.
Goo H
Korean J Radiol. 2021; 22(6):901-911.
PMID: 33660457
PMC: 8154780.
DOI: 10.3348/kjr.2020.0787.
Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake.
Pan Y, Zhou S, Wang Y, Li Q, Zhu T, Liu C
Curr Med Sci. 2020; 40(1):178-183.
PMID: 32166681
DOI: 10.1007/s11596-020-2162-8.
Carotid CTA at the Lowest Tube Voltage (70 kV) in Comparison with Automated Tube Voltage Adaption.
Eller A, Wiesmuller M, Wust W, Heiss R, Kopp M, Saake M
AJNR Am J Neuroradiol. 2019; 40(8):1374-1382.
PMID: 31296525
PMC: 7048475.
DOI: 10.3174/ajnr.A6108.
Application of Low Tube Potentials in CCTA: Results From the PROTECTION VI Study.
Stocker T, Leipsic J, Hadamitzky M, Chen M, Rubinshtein R, Deseive S
JACC Cardiovasc Imaging. 2019; 13(2 Pt 1):425-434.
PMID: 31202772
PMC: 8387300.
DOI: 10.1016/j.jcmg.2019.03.030.
Computed tomography portography of patients with cirrhosis with normal body mass index: Comparison between low-tube-voltage CT with low contrast agent dose and conventional CT.
Liu S, Sheng H, Shi H, Li W, Fan J, He J
Medicine (Baltimore). 2018; 97(48):e13141.
PMID: 30508890
PMC: 6283149.
DOI: 10.1097/MD.0000000000013141.
Contrast dose reduction with shortened injection durations in coronary CT angiography on 16-cm Wide-detector CT scanner.
Qu T, Li J, Jiao X, Zhang X, Song Z, Guo Y
Br J Radiol. 2018; 91(1092):20180580.
PMID: 30160183
PMC: 6319827.
DOI: 10.1259/bjr.20180580.
Dual-energy computed tomography using a gantry-based preclinical cone-beam microcomputed tomography scanner.
Tse J, Dunmore-Buyze J, Drangova M, Holdsworth D
J Med Imaging (Bellingham). 2018; 5(3):033503.
PMID: 30155511
PMC: 6103383.
DOI: 10.1117/1.JMI.5.3.033503.
Low tube voltage prospectively ECG-triggered coronary CT angiography: a systematic review of image quality and radiation dose.
Tan S, Yeong C, Aman R, Ng K, Abdul Aziz Y, Chee K
Br J Radiol. 2018; 91(1088):20170874.
PMID: 29493261
PMC: 6209486.
DOI: 10.1259/bjr.20170874.
German cardiac CT registry: indications, procedural data and clinical consequences in 7061 patients undergoing cardiac computed tomography.
Marwan M, Achenbach S, Korosoglou G, Schmermund A, Schneider S, Bruder O
Int J Cardiovasc Imaging. 2017; 34(5):807-819.
PMID: 29197025
DOI: 10.1007/s10554-017-1282-0.
Training and Service Provision in Cardiovascular CT: International Challenges and Solutions.
Nicol E, Shambrook J, Shaw E, Leipsic J, Shaw L
Curr Cardiol Rep. 2017; 19(12):123.
PMID: 29046971
DOI: 10.1007/s11886-017-0936-9.
Recent advances in cardiac computed tomography dose reduction strategies: a review of scientific evidence and technical developments.
Hedgire S, Baliyan V, Ghoshhajra B, Kalra M
J Med Imaging (Bellingham). 2017; 4(3):031211.
PMID: 28894760
PMC: 5571428.
DOI: 10.1117/1.JMI.4.3.031211.
Low-Dose Scanning Technology Combined with Low-Concentration Contrast Material in Renal Computed Tomography Angiography (CTA): A Preliminary Study.
Liu S, Li W, Shi H, Sheng H, Fan J, He J
Med Sci Monit. 2017; 23:4351-4359.
PMID: 28888094
PMC: 5602168.
DOI: 10.12659/msm.902917.