Intervendor Consistency and Reproducibility of Left Ventricular 2D Global and Regional Strain with Two Different High-end Ultrasound Systems

Abstract

Aims: We aimed to assess intervendor agreement of global (GLS) and regional longitudinal strain by vendor-specific software after EACVI/ASE Industry Task Force Standardization Initiatives for Deformation Imaging.

Methods And Results: Fifty-five patients underwent prospective dataset acquisitions on the same day by the same operator using two commercially available cardiac ultrasound systems (GE Vivid E9 and Philips iE33). GLS and regional peak longitudinal strain were analyzed offline using corresponding vendor-specific software (EchoPAC BT13 and QLAB version 10.3). Absolute mean GLS measurements were similar between the two vendors (GE -17.5 ± 5.2% vs. Philips -18.9 ± 5.1%, P = 0.15). There was excellent intervendor correlation of GLS by the same observer [r = 0.94, P < 0.0001; bias -1.3%, 95% CI limits of agreement (LOA) -4.8 to 2.2%). Intervendor comparison for regional longitudinal strain by coronary artery territories distribution were: LAD: r = 0.85, P < 0.0001; bias 0.5%, LOA -5.3 to 6.4%; RCA: r = 0.88, P < 0.0001; bias -2.4%, LOA -8.6 to 3.7%; LCX: r = 0.76, P < 0.0001; bias -5.3%, LOA -10.6 to 2.0%. Intervendor comparison for regional longitudinal strain by LV levels were: basal: r = 0.86, P < 0.0001; bias -3.6%, LOA -9.9 to 2.0%; mid: r = 0.90, P < 0.0001; bias -2.6%, LOA -7.8 to 2.6%; apical: r = 0.74; P < 0.0001; bias -1.3%, LOA -9.4 to 6.8%.

Conclusions: Intervendor agreement in GLS and regional strain measurements have significantly improved after the EACVI/ASE Task Force Strain Standardization Initiatives. However, significant wide LOA still exist, especially for regional strain measurements, which remains relevant when considering vendor-specific software for serial measurements.

Citing Articles

Deep learning improves test-retest reproducibility of regional strain in echocardiography.

Nyberg J, Ostvik A, Salte I, Olaisen S, Karlsen S, Dahlslett T Eur Heart J Imaging Methods Pract. 2024; 2(4):qyae092.

PMID: 39449961 PMC: 11498295. DOI: 10.1093/ehjimp/qyae092.


Influence of ultrasound transmit frequency on measurement of global longitudinal strain on 2D speckle tracking echocardiography.

Iwakura K, Onishi T, Koyama Y, Iwamoto M, Watanabe S, Tanaka K Sci Rep. 2023; 13(1):22196.

PMID: 38097659 PMC: 10721605. DOI: 10.1038/s41598-023-49664-3.


Progression of echocardiographic parameters and prognosis in transthyretin cardiac amyloidosis.

Chacko L, Karia N, Venneri L, Bandera F, Dal Passo B, Buonamici L Eur J Heart Fail. 2022; 24(9):1700-1712.

PMID: 35779241 PMC: 10108569. DOI: 10.1002/ejhf.2606.


Effect of tafamidis on global longitudinal strain and myocardial work in transthyretin cardiac amyloidosis.

Giblin G, Cuddy S, Gonzalez-Lopez E, Sewell A, Murphy A, Dorbala S Eur Heart J Cardiovasc Imaging. 2022; 23(8):1029-1039.

PMID: 35274130 PMC: 9303005. DOI: 10.1093/ehjci/jeac049.


Advantage and validation of vendor-independent software for myocardial strain analysis compared to vendor-specific software.

Chamberlain R, Shiino K, Scalia G, Sabapathy S, Chan J Australas J Ultrasound Med. 2021; 24(1):48-57.

PMID: 34760611 PMC: 8412014. DOI: 10.1002/ajum.12229.