Clinical Validation of Automated Audiometry with Continuous Noise-monitoring in a Clinically Heterogeneous Population Outside a Sound-treated Environment

Overview
Journal Int J Audiol
Publisher Informa Healthcare
Date 2016 May 22
PMID 27206551
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Examine the accuracy of automated audiometry in a clinically heterogeneous population of adults using the KUDUwave automated audiometer.

Design: Prospective accuracy study. Manual audiometry was performed in a sound-treated room and automated audiometry was not conducted in a sound-treated environment.

Study Sample: 42 consecutively recruited participants from a tertiary otolaryngology department in Western Australia.

Results: Absolute mean differences ranged between 5.12-9.68 dB (air-conduction) and 8.26-15 dB (bone-conduction). A total of 86.5% of manual and automated 4FAs were within 10 dB (i.e. ±5 dB); 94.8% were within 15 dB. However, there were significant (p < 0.05) differences between automated and manual audiometry at 250, 500, 1000, and 2000 Hz (air-conduction) and 500 and 1000 Hz (bone-conduction). The effect of age (≥55 years) on accuracy (p = 0.014) was not significant on linear regression (p > 0.05; R(2) =( ) 0.11). The presence of a hearing loss (better ear ≥26 dB) did not significantly affect accuracy (p = 0.604; air-conduction), (p = 0.218; bone-conduction).

Conclusions: This study provides clinical validation of automated audiometry using the KUDUwave in a clinically heterogeneous population, without the use of a sound-treated environment. Whilst threshold variations were statistically significant, future research is needed to ascertain the clinical significance of such variation.

Citing Articles

Accuracy and time efficiency of an automated commercial audiometer in the evaluation of hearing in quiet and noisy environments.

Yi Y, Wang X, Xiong Q, Xiong G Sci Prog. 2025; 108(1):368504251324348.

PMID: 40007331 PMC: 11863217. DOI: 10.1177/00368504251324348.


Prevalence of hearing loss and factors associated with hearing loss in Ethiopia: findings from the 2023 National Ethiopia Hearing Survey.

Meshesha A, Froschl U, Kebede M, Biratu T, Worku Y, Hunduma F BMJ Open. 2025; 15(1):e086288.

PMID: 39753268 PMC: 11883615. DOI: 10.1136/bmjopen-2024-086288.


Comparisons of air-conduction hearing thresholds between manual and automated methods in a commercial audiometer.

Liu H, Fu X, Li M, Wang S Front Neurosci. 2024; 17:1292395.

PMID: 38188027 PMC: 10771286. DOI: 10.3389/fnins.2023.1292395.


Clinical comparison of two automated audiometry procedures.

Liu H, Du B, Liu B, Fu X, Wang Y Front Neurosci. 2022; 16:1011016.

PMID: 36303947 PMC: 9595274. DOI: 10.3389/fnins.2022.1011016.


Validity of automated audiometry for hearing examination in patients with multidrug-resistant tuberculosis.

Purnami N, Palandeng R, Soedarsono S, Arifianto D, Moon I F1000Res. 2022; 10:1277.

PMID: 36127887 PMC: 9468620. DOI: 10.12688/f1000research.75090.2.