Clinical Validation of Automated Audiometry with Continuous Noise-monitoring in a Clinically Heterogeneous Population Outside a Sound-treated Environment
Overview
Authors
Affiliations
Objective: Examine the accuracy of automated audiometry in a clinically heterogeneous population of adults using the KUDUwave automated audiometer.
Design: Prospective accuracy study. Manual audiometry was performed in a sound-treated room and automated audiometry was not conducted in a sound-treated environment.
Study Sample: 42 consecutively recruited participants from a tertiary otolaryngology department in Western Australia.
Results: Absolute mean differences ranged between 5.12-9.68 dB (air-conduction) and 8.26-15 dB (bone-conduction). A total of 86.5% of manual and automated 4FAs were within 10 dB (i.e. ±5 dB); 94.8% were within 15 dB. However, there were significant (p < 0.05) differences between automated and manual audiometry at 250, 500, 1000, and 2000 Hz (air-conduction) and 500 and 1000 Hz (bone-conduction). The effect of age (≥55 years) on accuracy (p = 0.014) was not significant on linear regression (p > 0.05; R(2) =( ) 0.11). The presence of a hearing loss (better ear ≥26 dB) did not significantly affect accuracy (p = 0.604; air-conduction), (p = 0.218; bone-conduction).
Conclusions: This study provides clinical validation of automated audiometry using the KUDUwave in a clinically heterogeneous population, without the use of a sound-treated environment. Whilst threshold variations were statistically significant, future research is needed to ascertain the clinical significance of such variation.
Yi Y, Wang X, Xiong Q, Xiong G Sci Prog. 2025; 108(1):368504251324348.
PMID: 40007331 PMC: 11863217. DOI: 10.1177/00368504251324348.
Meshesha A, Froschl U, Kebede M, Biratu T, Worku Y, Hunduma F BMJ Open. 2025; 15(1):e086288.
PMID: 39753268 PMC: 11883615. DOI: 10.1136/bmjopen-2024-086288.
Liu H, Fu X, Li M, Wang S Front Neurosci. 2024; 17:1292395.
PMID: 38188027 PMC: 10771286. DOI: 10.3389/fnins.2023.1292395.
Clinical comparison of two automated audiometry procedures.
Liu H, Du B, Liu B, Fu X, Wang Y Front Neurosci. 2022; 16:1011016.
PMID: 36303947 PMC: 9595274. DOI: 10.3389/fnins.2022.1011016.
Purnami N, Palandeng R, Soedarsono S, Arifianto D, Moon I F1000Res. 2022; 10:1277.
PMID: 36127887 PMC: 9468620. DOI: 10.12688/f1000research.75090.2.