» Articles » PMID: 27160300

A Novel Alternative to Classify Tissues from T 1 and T 2 Relaxation Times for Prostate MRI

Overview
Journal MAGMA
Publisher Springer
Date 2016 May 11
PMID 27160300
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To segment and classify the different attenuation regions from MRI at the pelvis level using the T 1 and T 2 relaxation times and anatomical knowledge as a first step towards the creation of PET/MR attenuation maps.

Materials And Methods: Relaxation times were calculated by fitting the pixel-wise intensities of acquired T 1- and T 2-weighted images from eight men with inversion-recovery and multi-echo multi-slice spin-echo sequences. A decision binary tree based on relaxation times was implemented to segment and classify fat, muscle, prostate, and air (within the body). Connected component analysis and an anatomical knowledge-based procedure were implemented to localize the background and bone.

Results: Relaxation times at 3 T are reported for fat (T 1 = 385 ms, T 2 = 121 ms), muscle (T 1 = 1295 ms, T 2 = 40 ms), and prostate (T 1 = 1700 ms, T 2 = 80 ms). The relaxation times allowed the segmentation-classification of fat, prostate, muscle, and air, and combined with anatomical knowledge, they allowed classification of bone. The good segmentation-classification of prostate [mean Dice similarity score (mDSC) = 0.70] suggests a viable implementation in oncology and that of fat (mDSC = 0.99), muscle (mDSC = 0.99), and bone (mDSCs = 0.78) advocates for its implementation in PET/MR attenuation correction.

Conclusion: Our method allows the segmentation and classification of the attenuation-relevant structures required for the generation of the attenuation map of PET/MR systems in prostate imaging: air, background, bone, fat, muscle, and prostate.

Citing Articles

The value of synthetic magnetic resonance imaging in the diagnosis and assessment of prostate cancer aggressiveness.

Gao Z, Xu X, Sun H, Li T, Ding W, Duan Y Quant Imaging Med Surg. 2024; 14(8):5473-5489.

PMID: 39143997 PMC: 11320532. DOI: 10.21037/qims-24-291.


A review of PET attenuation correction methods for PET-MR.

Krokos G, MacKewn J, Dunn J, Marsden P EJNMMI Phys. 2023; 10(1):52.

PMID: 37695384 PMC: 10495310. DOI: 10.1186/s40658-023-00569-0.


The Preoperative Diagnostic Performance of Multi-Parametric Quantitative Assessment in Rectal Carcinoma: A Preliminary Study Using Synthetic Magnetic Resonance Imaging.

Zhu K, Chen Z, Cui L, Zhao J, Liu Y, Cao J Front Oncol. 2022; 12:682003.

PMID: 35707367 PMC: 9190242. DOI: 10.3389/fonc.2022.682003.


Synthetic MRI for Radiotherapy Planning for Brain and Prostate Cancers: Phantom Validation and Patient Evaluation.

Gouel P, Hapdey S, Dumouchel A, Gardin I, Torfeh E, Hinault P Front Oncol. 2022; 12:841761.

PMID: 35515105 PMC: 9065558. DOI: 10.3389/fonc.2022.841761.


Pseudo-T2 mapping for normalization of T2-weighted prostate MRI.

Sorland K, Sunoqrot M, Sandsmark E, Langorgen S, Bertilsson H, Trimble C MAGMA. 2022; 35(4):573-585.

PMID: 35150363 PMC: 9363383. DOI: 10.1007/s10334-022-01003-9.


References
1.
Zaidi H, Montandon M, Slosman D . Magnetic resonance imaging-guided attenuation and scatter corrections in three-dimensional brain positron emission tomography. Med Phys. 2003; 30(5):937-48. DOI: 10.1118/1.1569270. View

2.
Bezrukov I, Mantlik F, Schmidt H, Scholkopf B, Pichler B . MR-Based PET attenuation correction for PET/MR imaging. Semin Nucl Med. 2012; 43(1):45-59. DOI: 10.1053/j.semnuclmed.2012.08.002. View

3.
Sung K, Daniel B, Hargreaves B . Transmit B1+ field inhomogeneity and T1 estimation errors in breast DCE-MRI at 3 tesla. J Magn Reson Imaging. 2013; 38(2):454-9. PMC: 3620726. DOI: 10.1002/jmri.23996. View

4.
Deoni S . High-resolution T1 mapping of the brain at 3T with driven equilibrium single pulse observation of T1 with high-speed incorporation of RF field inhomogeneities (DESPOT1-HIFI). J Magn Reson Imaging. 2007; 26(4):1106-11. DOI: 10.1002/jmri.21130. View

5.
Edden R, Smith S, Barker P . Longitudinal and multi-echo transverse relaxation times of normal breast tissue at 3 Tesla. J Magn Reson Imaging. 2010; 32(4):982-7. PMC: 3154635. DOI: 10.1002/jmri.22306. View