» Articles » PMID: 27135144

A Novel Free Flap Monitoring System Using Tissue Oximetry with Text Message Alerts

Overview
Date 2016 May 3
PMID 27135144
Citations 9
Authors
Affiliations
Soon will be listed here.
Abstract

Background The time to detection of vascular compromise is a significant predictor of free flap salvage outcomes as early reexploration improves salvage rates. Continuous transcutaneous near-infrared tissue oximetry is an objective, quantitative method of detecting flap vascular compromise and has been shown to allow earlier reexploration and higher salvage rates than clinical assessment alone. We designed a novel text messaging system to improve communication using tissue oximetry monitoring. Methods A retrospective review was performed of a prospectively collected database of all microsurgical breast reconstructions from 2008 to 2015. A novel text messaging system was introduced in 2013 and programmed to send text messages alert when the tissue oximetry readings suggested potential flap compromise based on established thresholds. Patient demographics and complications, including rate of reexploration and flap loss were assessed. Results There were 900 autologous microsurgical breast free flaps during the study period: 614 were monitored with standard clinical monitoring and tissue oximetry compared with 286 flaps with the additional text messaging system. There were 27 unplanned returns to the operating room in the tissue oximetry group and 5 in the text messaging group with 1 complete flap loss in each group. Reexploration occurred sooner as a result of these text message alerts (17.5 vs. 26.6 hours postoperatively), however, it did not achieve statistical significance. Conclusions We were able to demonstrate the use of a novel text messaging system for tissue oximetry. This alert system shows promise in identifying impending flap loss with rapid notification of the surgical team.

Citing Articles

An Experimental and Clinical Study of Flap Monitoring with an Analysis of the Clinical Course of the Flap Using an Infrared Thermal Camera.

Kim H, Kwak S, Byeon J, Lee D, Kim J, Lim S Bioengineering (Basel). 2024; 11(7).

PMID: 39061770 PMC: 11273470. DOI: 10.3390/bioengineering11070688.


Development and usability of a Mobile Interactive Application (VCPW) for Vascular Crisis Prewarning after Skin Flap Transplantation.

Peng L, Xu L, Wang S, Huang W, Liu Q, Huang N JPRAS Open. 2023; 37:109-120.

PMID: 37520027 PMC: 10384608. DOI: 10.1016/j.jpra.2023.06.012.


Simple and Easily Interpretable Flap Monitoring Method Using a Commercial Pulse Oximeter and a Widely Used Bedside Patient Monitor.

Atomura D, Hoshino S, Osaki T, Sakakibara S Cureus. 2023; 14(12):e32549.

PMID: 36654599 PMC: 9840461. DOI: 10.7759/cureus.32549.


Flap Monitoring Using Transcutaneous Oxygen or Carbon Dioxide Measurements.

Halani S, Hembd A, Li X, Kirby B, Beard C, Haddock N J Hand Microsurg. 2022; 14(1):10-18.

PMID: 35256823 PMC: 8898164. DOI: 10.1055/s-0040-1718862.


Near-Infrared Spectroscopy (NIRS) versus Hyperspectral Imaging (HSI) to Detect Flap Failure in Reconstructive Surgery: A Systematic Review.

Lindelauf A, Saelmans A, van Kuijk S, van der Hulst R, Schols R Life (Basel). 2022; 12(1).

PMID: 35054458 PMC: 8778121. DOI: 10.3390/life12010065.