ICD Discrimination of SVT Versus VT with 1:1 V-A Conduction: A Review of the Literature

Overview
Date 2016 May 3
PMID 27134440
Citations 3
Authors
Affiliations
Soon will be listed here.
Abstract

Inappropriate ICD shocks are associated with increased mortality. They also impair patients' quality of life, increase hospitalizations, and raise health-care costs. Nearly 80% of inappropriate ICD shocks are caused by supraventricular tachycardia. Here we report the case of a patient who received a single-lead dual-chamber sensing ICD for primary prevention of sudden cardiac death and experienced inappropriate ICD shocks. V-A time, electrogram morphology, and response to antitachycardia pacing suggested atrioventricular nodal reentry tachycardia, which was confirmed in an electrophysiology study. Inspired by this case, we performed a literature review to discuss mechanisms for discrimination of supraventricular tachycardia with 1:1 A:V relationship from ventricular tachycardia with 1:1 retrograde conduction.

Citing Articles

Inappropriate implantable cardioverter-defibrillator shocks in repaired tetralogy of fallot patients: Prevalence and electrophysiological mechanisms.

Goldenthal I, Rosenbaum M, Lewis M, Sciacca R, Garan H, Biviano A Int J Cardiol Heart Vasc. 2020; 28:100543.

PMID: 32490149 PMC: 7256636. DOI: 10.1016/j.ijcha.2020.100543.


Repeated misclassifications of tachycardia by an implantable cardiac defibrillator.

Watanabe A, Fujimura O Indian Pacing Electrophysiol J. 2019; 20(1):35-38.

PMID: 31843557 PMC: 6994405. DOI: 10.1016/j.ipej.2019.12.004.


Atrioventricular 1:1 tachycardia effectively terminated by antitachycardia pacing : What is your diagnosis?.

Burri H Herzschrittmacherther Elektrophysiol. 2016; 27(3):244-8.

PMID: 27459976 DOI: 10.1007/s00399-016-0444-4.

References
1.
Singh D, Viswanathan M, Tanel R, Lee R, Lee B, Marcus G . His overdrive pacing during supraventricular tachycardia: a novel maneuver for distinguishing atrioventricular nodal reentrant tachycardia from atrioventricular reciprocating tachycardia. Heart Rhythm. 2014; 11(8):1327-35. DOI: 10.1016/j.hrthm.2014.04.038. View

2.
Biffi M . ICD programming. Indian Heart J. 2014; 66 Suppl 1:S88-100. PMC: 4237300. DOI: 10.1016/j.ihj.2013.11.007. View

3.
Connolly S, Gent M, Roberts R, Dorian P, Roy D, Sheldon R . Canadian implantable defibrillator study (CIDS) : a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation. 2000; 101(11):1297-302. DOI: 10.1161/01.cir.101.11.1297. View

4.
Daubert J, Zareba W, Cannom D, McNitt S, Rosero S, Wang P . Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact. J Am Coll Cardiol. 2008; 51(14):1357-65. DOI: 10.1016/j.jacc.2007.09.073. View

5.
STEVENSON W, Friedman P, Sager P, Saxon L, Kocovic D, Harada T . Exploring postinfarction reentrant ventricular tachycardia with entrainment mapping. J Am Coll Cardiol. 1997; 29(6):1180-9. DOI: 10.1016/s0735-1097(97)00065-x. View