Abstract

Background: Optimal adjustment of the filling volume of laparoscopic adjustable gastric banding is challenging and commonly performed empirically. Patients with band over-inflation and gastric obstruction arrive at the emergency department complaining of recurrent vomiting. In cases of gastric obstruction, intra-band pressure measurement may assist in determining the amount of fluid that should be removed from the band; however, our investigations have determined that intra-band pressure assessment need not play a role in the treatment of gastric band obstruction.

Methods: In patients coming to the emergency department with gastric band obstruction, we measured intra-band pressure at arrival and following stepped removal of fluid, comparing the initial pressure with post-deflation pressure and measuring the volume of fluid removed.

Results: Forty-eight patients participated in the study. Forty-five patients had a low-pressure/high-volume band. Their mean baseline pressure was 54.6 ± 22.3 mmHg. The mean volume of fluid removed from the band was 1.3 ± 0.8 ml. The mean post-deflation pressure was 22.5 ± 16.3 mmHg. Nearly 30 % of patients required as little as 0.5 ml of fluid removal, and 60 % of them were free of symptoms with removal of 1 ml.

Conclusions: Our results indicate that intra-band pressure measurement is of little value for determining the amount of fluid that should be removed for treatment of band obstruction. We suggest the removal of fluid in volumes of 0.5 ml until symptoms are relieved. Only in complicated cases, such as in patients having recurrent obstructions, should additional modalities be employed for further management guidance.

Citing Articles

Case Report of Patient Presenting in Shock from Band Penetration into Stomach After LAGB Surgery: Diagnosis by Emergency EGD After Misdiagnosis by Abdominal CT.

Cappell M, Mogrovejo E, Desai T Dig Dis Sci. 2016; 61(11):3366-3368.

PMID: 27572940 DOI: 10.1007/s10620-016-4287-y.

References
1.
Kodner C, Hartman D . Complications of adjustable gastric banding surgery for obesity. Am Fam Physician. 2014; 89(10):813-8. View

2.
Lechner W, Gadenstatter M, Ciovica R, Kirchmayr W, Schwab G . In vivo band manometry: a new access to band adjustment. Obes Surg. 2005; 15(10):1432-6. DOI: 10.1381/096089205774859399. View

3.
Livingston E . The incidence of bariatric surgery has plateaued in the U.S. Am J Surg. 2010; 200(3):378-85. PMC: 2923252. DOI: 10.1016/j.amjsurg.2009.11.007. View

4.
Burgerhart J, Aarts E, van de Meeberg P, Berends F, Siersema P, Smout A . Esophageal motor responses to increasing adjustment of an implanted gastric band. Neurogastroenterol Motil. 2013; 25(7):587-e461. DOI: 10.1111/nmo.12113. View

5.
Rauth T, Eckhauser A, LaFleur B, Melvin W, Holzman M . Intraband pressure measurements describe a pattern of weight loss for patients with adjustable gastric bands. J Am Coll Surg. 2008; 206(5):926-32. DOI: 10.1016/j.jamcollsurg.2007.12.037. View