» Articles » PMID: 27114250

Assessment of the Reliability of the Sysmex XE-5000 Analyzer to Detect Platelet Clumps

Overview
Journal Lab Med
Date 2016 Apr 27
PMID 27114250
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Platelet counts generated by automated analyzers on blood specimens that contain platelet clumps are often inaccurate and require verification by blood-smear review. In this study, we assessed the reliability of the Sysmex XE-5000 instrument to detect platelet clumps.

Method: We reviewed automated complete blood count (CBC) results and the findings of the microscopic review of corresponding blood smears of 600 blood specimens specifically selected from the routine laboratory workload. The sensitivity, specificity, efficiency, positive predictive value (PPV), and negative predictive value (NPV) of its 2 platelet-associated flags (abnormal platelet-size distribution [PAD] flag and platelet-clumps [CLP] flag) were determined.

Results: The respective values for the sensitivity, specificity, efficiency, and PPV were 42%, 83%, 63%, and 1% for the PAD flag and 57%, 99%, 78%, and 37% for the CLP flag. The NPV was 100%.

Conclusion: The overall reliability of the CLP flag is superior than that of the PAD flag but there is room for further improvement.

Citing Articles

The diagnostic accuracy of Sysmex XN for identification of pseudothrombocytopenia using various thresholds for definition of platelet aggregation.

Lunde H, Hjelmtvedt A, Amundsen E Int J Lab Hematol. 2022; 44(5):854-860.

PMID: 35751410 PMC: 9545299. DOI: 10.1111/ijlh.13920.


The flagging features of the Sysmex XN-10 analyser for detecting platelet clumps and the impacts of platelet clumps on complete blood count parameters.

Xu P, Fang K, Chen X, Liu Y, Dong Z, Zhu J Clin Chem Lab Med. 2022; 60(5):748-755.

PMID: 35212492 DOI: 10.1515/cclm-2021-1226.


Resistance to Acetylsalicylic Acid in Patients with Coronary Heart Disease Is the Result of Metabolic Activity of Platelets.

Grinshtein Y, Savchenko A, Kosinova A, Goncharov M Pharmaceuticals (Basel). 2020; 13(8).

PMID: 32752170 PMC: 7466119. DOI: 10.3390/ph13080178.


Peculiarities of Platelet Metabolism in Patients with Acute Coronary Syndrome with Anxiety-Depressive Disorders and Informativity of Enzymes in the Forecast of Development of Cardiovascular Complications.

Shimokhina N, Savchenko A, Petrova M Pharmaceuticals (Basel). 2020; 13(8).

PMID: 32731561 PMC: 7466177. DOI: 10.3390/ph13080169.

References
1.
Sandhaus L, Osei E, Agrawal N, Dillman C, Meyerson H . Platelet counting by the coulter LH 750, sysmex XE 2100, and advia 120: a comparative analysis using the RBC/platelet ratio reference method. Am J Clin Pathol. 2002; 118(2):235-41. DOI: 10.1309/MK3G-MC3V-P06R-PNV2. View

2.
Arneth B, Menschikowki M . Technology and new fluorescence flow cytometry parameters in hematological analyzers. J Clin Lab Anal. 2014; 29(3):175-83. PMC: 6807107. DOI: 10.1002/jcla.21747. View

3.
Gulati G, Uppal G, Florea A, Gong J . Detection of Platelet Clumps on Peripheral Blood Smears by CellaVision DM96 System and Microscopic Review. Lab Med. 2014; 45(4):368-71. DOI: 10.1309/LM604RQVKVLRFXOR. View