Quantification of the Effect of Pressure Wire Drift on the Diagnostic Performance of Fractional Flow Reserve, Instantaneous Wave-Free Ratio, and Whole-Cycle Pd/Pa

Overview
Date 2016 Apr 15
PMID 27076571
Citations 19
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Small drifts in intracoronary pressure measurements (±2 mm Hg) can affect stenosis categorization using pressure indices. This has not previously been assessed for fractional flow reserve (FFR), instantaneous wave-free ratio (iFR), and whole-cycle distal pressure/proximal pressure (Pd/Pa) indices.

Methods And Results: Four hundred forty-seven stenoses were assessed with FFR, iFR, and whole-cycle Pd/Pa. Cut point values for significance were predefined as ≤0.8, <0.90, and <0.93, respectively. Pressure wire drift was simulated by offsetting the distal coronary pressure trace by ±2 mm Hg. FFR, iFR, and whole-cycle Pd/Pa indices were recalculated and stenosis misclassification quantified. Median (±median absolute deviation) values for FFR, iFR, and whole-cycle Pd/Pa were 0.81 (±0.11), 0.90 (±0.07), and 0.93 (±0.06), respectively. For the cut point of FFR, iFR, and whole-cycle Pd/Pa, 34.6% (155), 50.1% (224), and 62.2% (278) of values, respectively, lay within ±0.05 U. With ±2 mm Hg pressure wire drift, 21% (94), 25% (110), and 33% (148) of the study population were misclassified with FFR, iFR, and whole-cycle Pd/Pa, respectively. Both FFR and iFR had significantly lower misclassification than whole-cycle Pd/Pa (P<0.001). There was no statistically significant difference between the diagnostic performance of FFR and iFR (P=0.125).

Conclusions: In a substantial proportion of cases, small amounts of pressure wire drift are enough to cause stenoses to change classification. Whole-cycle Pd/Pa is more vulnerable to such reclassification than FFR and iFR.

Citing Articles

Effect of prolonged pressure equalization on final drifting during pressure wire studies.

Jong C, Lu T, Lin L, Chen T, Liao M, Kuo J Sci Rep. 2024; 14(1):11504.

PMID: 38769360 PMC: 11106059. DOI: 10.1038/s41598-024-62440-1.


Assessment of the functional severity of coronary lesions from optical coherence tomography based on ensembled learning.

Tache I, Hatfaludi C, Puiu A, Itu L, Popa-Fotea N, Calmac L Biomed Eng Online. 2023; 22(1):127.

PMID: 38104144 PMC: 10724936. DOI: 10.1186/s12938-023-01192-x.


Coronary Physiology: Modern Concepts for the Guidance of Percutaneous Coronary Interventions and Medical Therapy.

Verdoia M, Rognoni A J Clin Med. 2023; 12(6).

PMID: 36983275 PMC: 10057250. DOI: 10.3390/jcm12062274.


Novel Method to Detect Pitfalls of Intracoronary Pressure Measurements by Pressure Waveform Analysis.

Jenei C, Tar B, Agoston A, Santa P, Santa J, Csippa B J Pers Med. 2022; 12(12).

PMID: 36556256 PMC: 9784866. DOI: 10.3390/jpm12122035.


Comparative analysis of instantaneous wave-free ratio and quantitative real-time myocardial contrast echocardiography for the assessment of myocardial perfusion.

Liang L, Zhu Y, Li F, Guo K, Chang S, Li Q Front Cardiovasc Med. 2022; 9:893647.

PMID: 36386338 PMC: 9643158. DOI: 10.3389/fcvm.2022.893647.


References
1.
Sen S, Asrress K, Nijjer S, Petraco R, Malik I, Foale R . Diagnostic classification of the instantaneous wave-free ratio is equivalent to fractional flow reserve and is not improved with adenosine administration. Results of CLARIFY (Classification Accuracy of Pressure-Only Ratios Against Indices Using Flow.... J Am Coll Cardiol. 2013; 61(13):1409-20. DOI: 10.1016/j.jacc.2013.01.034. View

2.
Nijjer S, Sen S, Petraco R, Escaned J, Echavarria-Pinto M, Broyd C . Pre-angioplasty instantaneous wave-free ratio pullback provides virtual intervention and predicts hemodynamic outcome for serial lesions and diffuse coronary artery disease. JACC Cardiovasc Interv. 2014; 7(12):1386-96. DOI: 10.1016/j.jcin.2014.06.015. View

3.
Eliasziw M, Donner A . Application of the McNemar test to non-independent matched pair data. Stat Med. 1991; 10(12):1981-91. DOI: 10.1002/sim.4780101211. View

4.
van de Hoef T, Nolte F, Damman P, Delewi R, Bax M, Chamuleau S . Diagnostic accuracy of combined intracoronary pressure and flow velocity information during baseline conditions: adenosine-free assessment of functional coronary lesion severity. Circ Cardiovasc Interv. 2012; 5(4):508-14. DOI: 10.1161/CIRCINTERVENTIONS.111.965707. View

5.
Spaan J, Piek J, Hoffman J, Siebes M . Physiological basis of clinically used coronary hemodynamic indices. Circulation. 2006; 113(3):446-55. DOI: 10.1161/CIRCULATIONAHA.105.587196. View